MPC1-1000 and MPC1-3000 User’s Manual
Page
25
© 2001 - 2007
DH Instruments, a Fluke Company
6.2.1
VARIABLE VOLUME SPECIFICATIONS
Table 1.
Variable Volume Specifications
MPC1-1000
MPC1-3000
Range
Vacuum to 1 000 psi (7 MPa)
Vacuum to 3 000 psi (21 MPa)
Proof
3 000 psi (21 MPa)
6 000 psi (42 MPa)
Burst
6 000 psi min (42 MPa)
12 000 psi min (84 MPa)
Total Volume
3.5 cu in. (57.5 cc)
2.5 cu in. (41 cc)
Revolutions for Total Volume
31 61