MPC1-1000 and MPC1-3000 User’s Manual
3.4.2
INCREASING PRESSURE
n
Close test outlet valve (4).
o
Rotate outlet selector valve (6) to vent position.
For small pressure changes, it may be possible to reach the desired pressure using the VV
only, proceed to Step
r
below.
p
Open pressure equalization valve (3) (pull valve outward).
OPENING THIS VALVE MAY RESULT IN A SUDDEN CHANGE IN THE TEST PRESSURE.
Ensure conditions are such that this either does not happen or that its effects
are inconsequential.
q
Place VV piston (2) at appropriate position, see Section 3.1.3.
r
Open test outlet valve until the approximate desired pressure is reached.
s
Close equalization valve (push valve inward).
t
Adjust VV knob until desired pressure is obtained (or until the piston floats).
u
To continue increasing pressures, repeat Step
p
or
t
shown above.
3.5
SETTING ABSOLUTE PRESSURE
3.5.1
INCREASING PRESSURE
n
Close test outlet valve (4).
o
Place outlet selection valve (6) to vacuum position.
p
Set pressure below atmosphere (if pressures below atmosphere are not required go to
next step).
q
Open pressure equalization valve (3) (pull valve outward).
r
Place VV piston (2) at appropriate position, see Section 3.1.3.
s
Open test outlet valve until the approximate desired pressure is reached.
t
Close equalization valve (push valve inward).
u
Adjust VV knob until desired pressure is obtained (or until the piston floats when
connected to a piston gauge).
© 2001 - 2007
DH Instruments, a Fluke Company
Page 10