MPC1-1000 and MPC1-3000 User’s Manual
5.4
PRESSURE CHANGES SLOWLY OR SLUGGISHLY
This is caused by a partially plugged filter or a large connected volume. Remove and clean or
replace the filter as needed.
The MPC1 uses very small inside diameter control tubes. The purpose of the filters is to protect these
tubes from contamination. Do NOT operate the MPC1 without the active filter elements of less than a
0.5 micron rating.
Pressure changes can be slow if a large external volume is connected to a MPC1. This condition is
compounded by a low supply pressure. If the pressure changes are too slow due to a large external
volume, reduce the volume.