1.1 PRODUCT 

OVERVIEW 

The purpose of this software is to calibrate the MFC analog control circuit in molbox1™.  There are three 
separate lines that are individually calibrated by the software.  First there is the set line that is used to set 
MFC set points.  Then there are two input lines, the sense and the measure.  The sense line is used  
to readjust the set value to match itself and the measure line is used to read the output of the MFC.   
The measure line is the most important of the three, even if the set point and sense point are out of 
tolerance, the measure line will still allow an accurate calibration if it is within tolerance.  In current (mA) 
mode there is no sense line so N/A appears in all sense fields while in current mode. 

1.2 GETTING 

STARTED 

Make sure that your PC has the following minimum requirements:  Windows

®

 3.0, 4 MB of RAM and 1 MB of 

hard disk space.  To install the software, insert the moltools installation disk and run the setup utility.  This can 
be done in Windows by going to the File Menu, selecting Run and typing “A:\SETUP”, or in Windows 95 
by pressing Start and choosing Run then typing “A:\SETUP” (where ‘A’ represents the drive that the 
installation disk is located).  This utility will install the molbox Analog Calibration program and all of its 
associated files into the moltools directory and create a program group in Windows named molbox 
moltools. 

Before attempting to run the Analog Calibration program, be sure that the molbox COM1 port is 
connected to one of the communications ports of the host PC with a standard RS232 cable.  This allows 
the program to auto-detect molbox and the serial communications setup is skipped.  To run the program 
double click the Analog Calibration Icon.  In most cases, if everything is properly attached, the software 
will auto-detect molbox and the calibration form will follow.  The exceptions are when the current settings 
of the molbox COM1 port are not Even parity, 1 Stop bit, 7 Data bits, and Baud rates of 2 400 or above.  
In these cases, you will have to manually set up the communications.  It is also possible that a general 
communications error will occur or some other unexpected problem that has not been mentioned.  If so, 
choose the Communications menu option and select the proper settings from the Serial Communications 
Form shown at the end of this document in the Forms/Diagrams section.  If you cannot determine the 
current settings of molbox, use the following procedure from the molbox keypad: 

 Press 

[ESC]

 until you return to the main menu if you are NOT already there. 

 

Select 

<0 Special>

, followed by 

<5 Internal>

 and 

<2 Remote>

.  Old style molbox1 users should 

select:  

<3 Special>

<8 Internal>

<2 Remote>

<1 com1>

 

 

If the molbox has an IEEE 488 option, select COM1 after selecting Remote in the previous step. 

 

Either change the baud rate with the number keys and select 

[ENTER]

, or bypass it to view the next 

parameter by pressing enter. 

 Repeat 

Step  for Parity, Word Length and Stop Bits, and terminating characters. 

  Select the proper configuration on the Serial Communications Form and press 

[OK]

Summary of Contents for MOLBOX ANALOG CIRCUIT

Page 1: ...2002 DH Instruments Inc molbox Analog Circuit Calibration Software User s Manual...

Page 2: ...wever no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual...

Page 3: ...4 3 MENU OPTIONS 5 3 1 FILE SAVE DATA SHORTCUT CTRL S 5 3 2 FILE PRINT FILE 5 3 3 FILE PRINT CURRENT DATA SHORTCUT CTRL P 5 3 4 FILE VIEW FILE SHORTCUT CTRL F 5 3 5 TEST CHECK MOLBOX CALIBRATION 5 3...

Page 4: ...Figure 3 Set Point Test Screen 9 Figure 4 molbox Analog Calibration Cable 11 Figure 5 Old Style molbox Analog Calibration Cable 11 A AB BO OU UT T T TH HI IS S M MA AN NU UA AL L Manual Conventions C...

Page 5: ...m group in Windows named molbox moltools Before attempting to run the Analog Calibration program be sure that the molbox COM1 port is connected to one of the communications ports of the host PC with a...

Page 6: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 2 N NO OT TE ES S...

Page 7: ...ation is out of specification This is accomplished by selecting a Voltage or Current mode Calibration Check from the Test menu see Section 3 5 The software will set two points 5 and 0 5 V or 20 and 5...

Page 8: ...ration option displays see Section 3 1 If for some reason the calibration fails the original calibration values are restored While the calibration is running it is not necessary to do anything except...

Page 9: ...FILE PRINT FILE This option allows the printing of a selected calibration file to the default Windows printer 3 3 FILE PRINT CURRENT DATA SHORTCUT CTRL P Print Current Data will print the current con...

Page 10: ...utput see Section 5 3 To set a value choose the desired set mode and enter the desired value and press either enter or send then enter the multimeter reading when prompted The percentage of full scale...

Page 11: ...ing to select any options check to make sure that everything is still plugged in properly To change the settings either click the settings that you want or tab to the field and use the arrow keys to c...

Page 12: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 8 N NO OT TE ES S...

Page 13: ...RM MS S D DI IA AG GR RA AM MS S 5 1 SERIAL COMMUNICATIONS FORM Figure 1 Communications Setup Screen 5 2 ANALOG CALIBRATION FORM Figure 2 molbox Analog Calibration Software ver 1 2b DH Instruments 19...

Page 14: ...25 lines including help text nor should there be any other colons other than those below Analog uses them as a delimiter The order and spacing of the following lines should also be the same Current U...

Page 15: ...5 FORMS DIAGRAMS Page 11 1997 2002 DH Instruments Inc 5 5 CABLES Figure 4 molbox Analog Calibration Cable Figure 5 Old Style molbox Analog Calibration Cable...

Page 16: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 12 N NO OT TE ES S...

Page 17: ...or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warran...

Page 18: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 14 N NO OT TE ES S...

Reviews: