background image

MFC SWITCHBOX USER’S MANUAL 

© 1997-2002 DH Instruments, Inc. 

Page 4

 

2.3 RECALIBRATE 

A recalibration determines new calibration coefficients for the set, measure and sense lines based on the 
supplied multimeter readings.  To perform a recalibration, choose either a Voltage or Current mode 
Recalibration from the Test Menu (see Section 3.6).  The software will set 5 and 0.5 V or 20 and 5 mA 
depending on the type of calibration selected.  After each set point the multimeter readings must be 
entered when prompted in the same manner as a calibration check, except it will do it twice.  The first 
time the software will calculate new adders and multipliers and write them to molbox1.  The second time, 
the software is attempting to verify that the new calibration was successful.  Therefore, the second set of 
data input by the user should be within the molbox specification.  Once the calibration passes, the save 
calibration option displays (see Section 3.1).  If for some reason the calibration fails, the original 
calibration values are restored. 

While the calibration is running it is not necessary to do anything except wait for the software to inform 
you when to input information or if there is a problem.  To stop any calibration early just press ESC.  If 
early, the original calibration values are written back to the molbox. 

There is a file named (molbox serial number (+_mA)).cfg that resides in the installation directory of the 
Analog program.  The file name will have ‘_mA’ after the serial number for current (mA) configuration files 
only.  In this file the current, previous, former and initial molbox calibration values and their dates are 
maintained.  In this case former refers to the value prior to previous.   

Data for this file is used to restore previous calibration values in order to maintain metrological continuity 
(see Section 3.6).  At the beginning and end of each calibration, the molbox calibration values are 
compared to the file data.  If they do not match the file is updated.  See Section 5.4 for an example .cfg file. 

 

Summary of Contents for MOLBOX ANALOG CIRCUIT

Page 1: ...2002 DH Instruments Inc molbox Analog Circuit Calibration Software User s Manual...

Page 2: ...wever no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual...

Page 3: ...4 3 MENU OPTIONS 5 3 1 FILE SAVE DATA SHORTCUT CTRL S 5 3 2 FILE PRINT FILE 5 3 3 FILE PRINT CURRENT DATA SHORTCUT CTRL P 5 3 4 FILE VIEW FILE SHORTCUT CTRL F 5 3 5 TEST CHECK MOLBOX CALIBRATION 5 3...

Page 4: ...Figure 3 Set Point Test Screen 9 Figure 4 molbox Analog Calibration Cable 11 Figure 5 Old Style molbox Analog Calibration Cable 11 A AB BO OU UT T T TH HI IS S M MA AN NU UA AL L Manual Conventions C...

Page 5: ...m group in Windows named molbox moltools Before attempting to run the Analog Calibration program be sure that the molbox COM1 port is connected to one of the communications ports of the host PC with a...

Page 6: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 2 N NO OT TE ES S...

Page 7: ...ation is out of specification This is accomplished by selecting a Voltage or Current mode Calibration Check from the Test menu see Section 3 5 The software will set two points 5 and 0 5 V or 20 and 5...

Page 8: ...ration option displays see Section 3 1 If for some reason the calibration fails the original calibration values are restored While the calibration is running it is not necessary to do anything except...

Page 9: ...FILE PRINT FILE This option allows the printing of a selected calibration file to the default Windows printer 3 3 FILE PRINT CURRENT DATA SHORTCUT CTRL P Print Current Data will print the current con...

Page 10: ...utput see Section 5 3 To set a value choose the desired set mode and enter the desired value and press either enter or send then enter the multimeter reading when prompted The percentage of full scale...

Page 11: ...ing to select any options check to make sure that everything is still plugged in properly To change the settings either click the settings that you want or tab to the field and use the arrow keys to c...

Page 12: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 8 N NO OT TE ES S...

Page 13: ...RM MS S D DI IA AG GR RA AM MS S 5 1 SERIAL COMMUNICATIONS FORM Figure 1 Communications Setup Screen 5 2 ANALOG CALIBRATION FORM Figure 2 molbox Analog Calibration Software ver 1 2b DH Instruments 19...

Page 14: ...25 lines including help text nor should there be any other colons other than those below Analog uses them as a delimiter The order and spacing of the following lines should also be the same Current U...

Page 15: ...5 FORMS DIAGRAMS Page 11 1997 2002 DH Instruments Inc 5 5 CABLES Figure 4 molbox Analog Calibration Cable Figure 5 Old Style molbox Analog Calibration Cable...

Page 16: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 12 N NO OT TE ES S...

Page 17: ...or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warran...

Page 18: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 14 N NO OT TE ES S...

Reviews: