2.1 OVERVIEW 

Before running a calibration, connect the 25 pin connector end of the calibration cable to the MFC port and the 
other end to the voltage or current measuring channel of the multimeter.  The red plug is positive voltage, 
the black plug is negative voltage, the blue plug is positive current and the white plug negative current. 

 

 

When performing a Voltage mode calibration only connect the Red and Black connectors to the 
multimeter.  Current (mA) calibrations should only have the blue and white connectors attached. 

 

 

Leaving different mode connectors attached or letting any connectors come in contact during a 
calibration could damage the molbox circuit. 

 

2.2 CHECK 

CALIBRATION 

The first step is to check to see if the calibration is out of specification.  This is accomplished by selecting 
a Voltage or Current mode Calibration Check from the Test menu (see Section 3.5).  The software will set 
two points, 5 and 0.5 V or 20 and 5 mA, without correcting with the molbox sense line.  After each setting 
a prompt on the status bar to enter the current multimeter reading displays.  The area of input will 
highlight and you must type in the multimeter reading and press return.  The multimeter will only display 
the proper value when you are prompted; so wait for the prompt before entering the reading. 

 

 

It is important to make sure that the value is properly input into the text box as this will directly 
affect the results of the calibration. 

 

The agreement between the multimeter readings and those read back from the molbox must be less than 
± 0.1 % full scale for set values and ± 0.05 % full scale for the measure values for molbox1.  If this is the 
case, the molbox1 is in tolerance and there are no other necessary steps.  However, if the molbox is out 
of tolerance then a recalibration is necessary. 

Summary of Contents for MOLBOX ANALOG CIRCUIT

Page 1: ...2002 DH Instruments Inc molbox Analog Circuit Calibration Software User s Manual...

Page 2: ...wever no warranty expressed or implied is provided DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual...

Page 3: ...4 3 MENU OPTIONS 5 3 1 FILE SAVE DATA SHORTCUT CTRL S 5 3 2 FILE PRINT FILE 5 3 3 FILE PRINT CURRENT DATA SHORTCUT CTRL P 5 3 4 FILE VIEW FILE SHORTCUT CTRL F 5 3 5 TEST CHECK MOLBOX CALIBRATION 5 3...

Page 4: ...Figure 3 Set Point Test Screen 9 Figure 4 molbox Analog Calibration Cable 11 Figure 5 Old Style molbox Analog Calibration Cable 11 A AB BO OU UT T T TH HI IS S M MA AN NU UA AL L Manual Conventions C...

Page 5: ...m group in Windows named molbox moltools Before attempting to run the Analog Calibration program be sure that the molbox COM1 port is connected to one of the communications ports of the host PC with a...

Page 6: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 2 N NO OT TE ES S...

Page 7: ...ation is out of specification This is accomplished by selecting a Voltage or Current mode Calibration Check from the Test menu see Section 3 5 The software will set two points 5 and 0 5 V or 20 and 5...

Page 8: ...ration option displays see Section 3 1 If for some reason the calibration fails the original calibration values are restored While the calibration is running it is not necessary to do anything except...

Page 9: ...FILE PRINT FILE This option allows the printing of a selected calibration file to the default Windows printer 3 3 FILE PRINT CURRENT DATA SHORTCUT CTRL P Print Current Data will print the current con...

Page 10: ...utput see Section 5 3 To set a value choose the desired set mode and enter the desired value and press either enter or send then enter the multimeter reading when prompted The percentage of full scale...

Page 11: ...ing to select any options check to make sure that everything is still plugged in properly To change the settings either click the settings that you want or tab to the field and use the arrow keys to c...

Page 12: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 8 N NO OT TE ES S...

Page 13: ...RM MS S D DI IA AG GR RA AM MS S 5 1 SERIAL COMMUNICATIONS FORM Figure 1 Communications Setup Screen 5 2 ANALOG CALIBRATION FORM Figure 2 molbox Analog Calibration Software ver 1 2b DH Instruments 19...

Page 14: ...25 lines including help text nor should there be any other colons other than those below Analog uses them as a delimiter The order and spacing of the following lines should also be the same Current U...

Page 15: ...5 FORMS DIAGRAMS Page 11 1997 2002 DH Instruments Inc 5 5 CABLES Figure 4 molbox Analog Calibration Cable Figure 5 Old Style molbox Analog Calibration Cable...

Page 16: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 12 N NO OT TE ES S...

Page 17: ...or goodwill even if DHI has been advised of the possibility thereof and regardless of whether such products are used individually or as components in other products Items returned to DHI under warran...

Page 18: ...MFC SWITCHBOX USER S MANUAL 1997 2002 DH Instruments Inc Page 14 N NO OT TE ES S...

Reviews: