DFI MDP156 User Manual Download Page 15

15

Chapter 4

 INDICATIONS FOR USE

User's Manual | 

MDP156

X

Precautions modifications

No modification of this equipment is allowed.

Do not modify this equipment without authorization of the manufacturer

If this equipment is modified, appropriate inspection and testing must be conducted to 

ensure continued safe use of the equipment.

If there is any other problem, please consult or return the device to your distributor. Do not 

try to repair a defective device.

X

Recommendation to use more than one unit

As problems, can occasionally occur, when the monitor is used under critical conditions, we 

strongly recommend you use more than one unit or prepare a spare unit for replacement.

X

About the LCD Display Panel

The LCD panel fitted to this unit is manufactured with high precision technology, giving a 

functioning pixel ratio of at least 99.99%. Thus, a very small proportion of pixels may be 

“stuck”, either always off (black), always on (red, green, or blue), or flashing. In addition, 

over a long period of use, because of the physical characteristics of the liquid crystal dis-

play, such “stuck” pixels may appear spontaneously. These problems are not a malfunc-

tion.

Do not leave the LCD screen facing the sun as it can damage the LCD screen. Take care 

when you place the unit by a window.

Do not store the monitor FACE DOWN.

Do not push or scratch the LCD screen.

Do not place a heavy object on the LCD screen. This may cause the screen to lose unifor-

mity.

If the unit is used in a cold place, a residual image may appear on the screen. This is not a 

malfunction. When the monitor becomes warm, the screen returns to normal.

The screen and the cabinet become warm during operation. This is not a malfunction.

X

Images that may cause burn-in

Masked / boarded images with aspect ratios other than 16:9

Color bars or images that remain static for a long time

Continuous characters or messages displaying on the screen

X

To reduce the risk of burn-in

Turn off the character displays from connected equipment.

X

About the screen protect panel

The screen protect panel is made of toughened glass, but there is a possibility that it may 

crack.

Handle with care. Avoid strong impact, such as dropping from a high place or an object 

swinging into it.

Do not scratch the panel with a sharp object or place it in harm’s way.

X

A long period of use

Due to the characteristics of LCD panel, displaying static images for extended periods, or 

using the unit repeatedly in a high temperature/high humidity environments may cause 

image smearing, burn-in, areas of which brightness is permanently changed, lines, or a 

decrease in overall brightness.

In particular, continued displaying of an image smaller than the monitor screen, such as in 

a different aspect ratio, may shorten the life of the unit.

Avoid displaying a still image for an extended period, or using the unit repeatedly in a high 

temperature/high humidity environment such an airtight room, or around the outlet of an 

air conditioner.

X

Moisture condensation

If the unit is suddenly taken from a cold to a warm location, or if ambient temperature sud-

denly rises, moisture may form on the outer surface of the unit and/or inside of the unit. This 

is known as condensation. If condensation occurs, turn off the unit and wait until the conden-

sation clears before operating the unit. Operating the unit while condensation is present may 

damage the unit.

X

Before cleaning

Be sure to disconnect the AC power cord from the AC outlet.

X

Cleaning the monitor

A material that withstands disinfection is used for the front protection plate of the healthcare 

use LCD monitor. The protection plate surface is specially treated to reduce reflection of light. 

When solvents such as benzene or thinner, or acid, alkaline or abrasive detergent or chemical 

cleaning cloth are used for the protection plate surface/monitor surface, the performance of 

the monitor may be impaired or the finish of the surface may be damaged. Take care with re-

spect to the following:

Clean the protection plate surface/monitor surface with a 50 to 70 v/v% concentration of 

isopropyl alcohol or a 76.9 to 81.4 v/v% concentration of ethanol using a swab method. 

Wipe the protection plate surface gently (wipe using less than 1 N force).

Stubborn stains may be removed with a soft cloth such as a cleaning cloth lightly damp-

ened with mild detergent solution using a swab method and then clean using the above 

chemical solution.

Never use solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or 

chemical cleaning cloth for cleaning or disinfection, as they will damage the protection 

plate surface/monitor surface.

Do not use unnecessary force to rub the protection plate surface/monitor surface with a 

stained cloth. The protection plate surface/monitor surface may be scratched.

Do not keep the protection plate surface/monitor surface in contact with a rubber or vinyl 

resin product for a long period of time. The finish of the surface may deteriorate.

Summary of Contents for MDP156

Page 1: ...MDP156 15 6 Healthcare Display User s Manual A 605 M 2034...

Page 2: ...CC and DOC Statement on Class B This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC rules These limits are designed to provid...

Page 3: ...notices for use in the healthcare applications 11 Guidance and manufacturer s declaration electromagnetic emissions 12 Guidance and manufacturer s declaration electromagnetic immunity 12 Recommended s...

Page 4: ...devices even before installing them in your system unit Static electrical discharge can damage computer components without causing any signs of physical damage You must take extra care in han dling t...

Page 5: ...dapter Power Cord VGA Cable HDMI Cable USB Cable for Touch Optional Items Cable Clamp x 3 pcs VESA Desk Stand The main body and accessories in the package may not come similar to the information liste...

Page 6: ...operation Mechanical Specification Mounting VESA Mount 75 x 75 mm 100 x 100 mm 200 x 100 mm VESA Screws M4x4L 75x75mm 100 x 100mm flat head Dimensions W x H x D 387 8 x 232 9 x 38 45 mm Housing Back c...

Page 7: ...to be played through the monitor speak ers 4 HDMI Transmits and protects copyrighted digital video and audio Example An HD tuner to an HD ready TV 5 VGA RGB Transmits video from a PC to a monitor Exam...

Page 8: ...ews to fix the desk stand to VESA holes on the back cover of the device Important The picture is for demonstration purposes only VESA Mount accessories are not supplied by default Connecting Periphera...

Page 9: ...To turn off the device please disconnect the power cord from an electric outlet Important To fully disengage the power please disconnect the input plug from the ac outlet To avoid risk of electric sh...

Page 10: ...cles trains high volt age lines or fluorescent lamps Bad color Color density or tint control may not be adjusted properly See the adjusted value of image Image appears and disappears Input may not be...

Page 11: ...n shall be addressed to the authorized representative For any service or guarantee matters please refer to the ad dresses provided in the separate service or guarantee documents X Important safeguards...

Page 12: ...c dis charge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air Floors should be wood concrete or ce ramic tile If floors are covered with synthetic material the relative humidity should...

Page 13: ...by an electromagnetic site survey A should be less than the compli ance level in each frequency range B Interference may occur near equipment marked with following symbol NOTE 1 At 80 MHz and 800 MHz...

Page 14: ...the patient to come into contact at the same time Grounding reliability can only be achieved when the equipment is connected to an equiva lent receptacle marked Hospital Only or Hospital Grade X Insta...

Page 15: ...ages for extended periods or using the unit repeatedly in a high temperature high humidity environments may cause image smearing burn in areas of which brightness is permanently changed lines or a dec...

Page 16: ...each single failed Product that is returned for servicing repair reworking or failure investigation to Seller or to the Authorized Service Provider At least one adhesive yellow label shall be attached...

Reviews: