12
Chapter 4
INDICATIONS FOR USE
User's Manual |
MDP156
X
Guidance and manufacturer’s declaration-electromagnetic emissions
This product is intended for use in the electromagnetic environment specified below.
The customer or the user should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The product uses RF energy only for its
internal function. Therefore, its RF emis-
sions are very low and are not likely
to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Class B
The product is suitable for use in all
establishments, including domestic
establishments and those directly con-
nected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
X
Guidance and manufacturer’s declaration - electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The cus-
tomer should assure that it is used in such as environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment-
guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be
wood, concrete or ce-
ramic tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical fast tran-
sient/burst
IEC 61000-4-4
±2 kV for power sup-
ply lines
±1 kV for input/ out-
put lines
±2 kV for power sup-
ply lines
±1 kV for input/ out-
put lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV differential
mode
±2 kV common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5%
(> 95% dip in) for 0.5
cycle
40%
(60% dip in) for 5
cycles
70%
(30% dip in) for 25
cycles
< 5%
(> 95% dip in) for 5
sec
< 5%
(> 95% dip in) for 0.5
cycle
40%
(60% dip in) for 5
cycles
70%
(30% dip in) for 25
cycles
< 5%
(> 95% dip in) for 5
sec
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment. If the user of
the product requires
continued operation
during power mains
interruptions, it is
recommended that
the product powered
from an uninterrupt-
ible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital