17
Analysis of Various Coatings Using Program G
Data from the ASTM Inter-Laboratory Study of the Calcium Hydride Method D7358 comparing the HydroSCOUT
results with Karl Fischer. The coatings analyzed were:
Fluid A: Clear Acrylic sealer
Fluid B: Acrylic white tint base
Fluid C: Acrylic stain
Fluid D: Phenolic resin containing approximately 10% unreacted Phenol.
Fluid E: Pigmented polyester polyol resin system containing MEK, Toluene, Ethyl Acetate, PM Acetate, and
Ethyl 3-Ethoxy Propionate.
Fluid F: Phenolic coating containing approximately 5% unreacted Phenol as well as Ethyl Alcohol, PM
Acetate, and Dipropylene Glycol Monomethyl Ether.
y = 0.9621x + 0.2562
R² = 0.9806
0
10
20
30
40
50
60
70
80
90
100
0
20
40
60
80
100
W
at
er
by
Hy
droSCOU
T
(%
)
Water by Karl Fischer (%)
Calcium Hydride vs Karl Fischer