Dexsil HydroSCOUT User Manual Download Page 10

 

4. 

ANALYSIS:

  Press  the  <ON>  key  and  make  sure  the  meter  is  set  for  Program  “A.”  (NOTE:  If  changing  the 

program, see “Selecting a Program” in the manual).  Once the meter displays “

tUbE

”, insert the reaction tube 

into  the  opening  at  the  bottom  of  the  meter.    The  hole  in  the  white  cap  must  be  lined  up  with  the  metal  pin 
extending from the inside of the hole of the meter.  Push the reaction tube all the way into the meter. Press the 
<READ>  key  on  the  face  of  the  meter.    The  display  will  flash  the  word  “

CALC

,”  and  the  results  will  be 

displayed  in  volume  percent  (v/v%)      Record  the  results.   

KEEP  THE  METER  AND  REACTION  TUBE 

UPRIGHT AT ALL TIMES.

  

 
5. 

DISPOSAL:

  Carefully  remove  the  reaction  tube  from  the  meter.    Vent  the  reaction  tube  by  inserting  it 

completely into the specially designed venting cap unit and remove. 

REACTION TUBE AND CAP MUST BE 

UPRIGHT  AND  POINTED  AWAY  FROM  USER  AND  BYSTANDERS.

    Tap  the  reaction  tube  on  a  hard 

surface to shake any loose glass to the bottom of the tube, then break the top (green) ampule in the reaction tube 
by compressing the sides of the reaction tube.  Shake the tube for 15 seconds.  After 5 minutes, vent the tube 
again.  Discard the reaction tube as laboratory waste. 

DO NOT REMOVE THE WHITE CAP.

 

 
 

Program B

 

– Procedure for Testing Water Content up to 100% 

 

(Reagent Systems: HS-ORP + HS-DVP-12)

 

 
1. 

PREPARATION:

    Remove  a  screw-capped  dilution  vial  from  the  fluted  cardboard  carton  and  a  sampling 

syringe and a white-capped reaction tube from the reagent box.  Check the contents to ensure that all items are 
present and intact.  Remove the white cap from the reaction tube and discard.  Carefully remove a green water 
ampule from the cardboard carton and insert it fully into the top of the  protective ampule sleeve containing the 
grayish white ampule.  Return the protective ampule sleeve to the reaction tube with the grayish white ampule 
on the bottom and place the reaction tube back into the reagent box. 

 
2. 

SAMPLE  INTRODUCTION:

  Work  the  plunger  of  the  empty  sampling  syringe  a  few  times  to  ensure  that  it 

slides  easily.    Place  the  tip  of  the  syringe  into  the  oil  sample  to  be  tested  (sample  must  be  well  mixed  and 
representative) and slowly pull back on the plunger until it reaches the stop and cannot be pulled further.  When 
drawing up the sample, make sure there are no air bubbles in the syringe.  If air bubbles are present, depress the 
plunger again and draw a second sample slowly.  Remove the syringe from the oil sample and wipe any excess 
oil from the outside of the syringe with a tissue wipe, be sure not to blot the tip of the syringe. 

 
3. 

SAMPLE  DILUTION: 

  Remove  the  screw-cap  from  the  dilution  vial.    Insert  the  sampling  syringe  into  the 

dilution vial and dispense the oil sample by depressing the plunger.  Proceed to mix the solution thoroughly by 
repeated drawing and re-dispensing of the solution into the dilution vial with the syringe (3 or 4 times).  Using 
the  same  sampling  syringe,  draw  up  a  sample  from  the  diluted  solution  wiping  any  excess  liquid  from  the 
outside  of  the  syringe.    Dispense  the  sample  into  the  reaction  tube.    Insert  the  black  septum  (hole  on  top)  by 
gripping the reaction tube firmly and pressing down on the septum with your thumb until flush with top of the 
tube.  

TIGHTLY

 cap the tube with the holed, white cap

.  NOTE: DO NOT SQUEEZE THE SIDES OF THE 

TUBE WHILE INSERTING THE BLACK SEPTUM.  DOING THIS WILL RESULT IN LOW READINGS.

 

 
4. 

REACTION:

    Break  the  bottom  (grayish  white)  ampule  in  the  reaction  tube  by  compressing  the  sides  of  the 

tube.  Mix thoroughly by shaking the reaction tube vigorously for 30 seconds.  Allow the reaction to proceed for 
a total of 2 minutes.  

DO NOT BREAK THE GREEN AMPULE.

 

 
5. 

ANALYSIS:

    Press  the  <ON>  key  and  make  sure  the  meter  is  set  for  Program  “

  B

” (NOTE: If changing the 

program, see “Selecting a Program” in the manual).  Once the meter displays “

tUbE

”, insert the reaction tube 

into  the  opening  at  the  bottom  of  the  meter.    The  hole  in  the  white  cap  must  be  lined  up  with  the  metal  pin 
extending from the inside of the hole of the meter.  Push the reaction tube all the way into the meter. Press the 

<READ>  key  on  the  face  of  the  meter.    The  display  will  flash  the  word  “

CALC

,”  and  the  results  will  be 

displayed  in  volume  percent  (v/v%).    Record  the  results.   

KEEP  THE  METER  AND  REACTION  TUBE 

UPRIGHT AT ALL TIMES.

 

 
6. 

DISPOSAL:

    Carefully  remove  the  reaction  tube  from  the  meter.    Vent  the  reaction  tube  by  inserting  it 

completely into the specially designed venting cap unit and remove.  

REACTION TUBE AND CAP MUST BE 

UPRIGHT  AND  POINTED  AWAY  FROM  USER  AND  BYSTANDERS

.    Tap  the  reaction  tube  on  a  hard 

surface to shake any loose glass to the bottom of the tube, then break the top (green) ampule in the reaction tube 
by compressing the sides of the reaction tube.  Shake the tube for 15 seconds.  After 5 minutes, vent the tube 
again.  Discard the reaction tube and dilution vial as laboratory waste.

 DO NOT REMOVE WHITE CAP.

 

 

Summary of Contents for HydroSCOUT

Page 1: ...aims any implied warranties of merchantability or fitness for any particular purpose Further Dexsil Corporation reserves the right to revise this publication and to make changes from time to time in t...

Page 2: ...BROKEN GLASS AS IT CAN PENETRATE THE PLASTIC TUBE AND CUT FINGERS THE GRAY AMPULE IN THE TUBES CONTAINS CALCIUM HYDRIDE WHICH IS A FLAMMABLE SOLID AND IS WATER REACTIVE IN CASE OF ACCIDENTAL BREAKAGE...

Page 3: ...gram A Procedure for Testing Water Content up to 20 6 7 Program B Procedure for Testing Water Content up to 100 7 Program C D E Procedure for Low Levels of Water 5 mL Sample Size 8 Program F Procedure...

Page 4: ...ogram is described below Table 1 HydroSCOUT Analysis Program Information Program Sample Matrix Types Tested1 Units MDL3 Maximum Reagents Needed Sample Size A Used Oil Liquid Used Oil Chemtool 250 Pain...

Page 5: ...the pressure function is a linear parameter the meter response value can be corrected to produce accurate results on all oil types The HydroSCOUT response on low water level oils has been documented...

Page 6: ...thod D 7358 In order to achieve a result based on the weight of the sample the sample must be weighed into the reaction vessel Not only does this allow for a weight based result but it improves accura...

Page 7: ...to G followed by OFF and then back to A 0 20 again Press the READ key at any point to fix the selection at the displayed program or to turn the meter off when OFF is displayed To begin using the selec...

Page 8: ...i e programs C D or E ethylene glycol and acids have been found to produce a positive result Ethylene glycol responds at approximately 7 of the volumetric equivalent Error Code Explanation Likely Caus...

Page 9: ...ule from the cardboard carton and insert it fully into the top of the protective ampule sleeve containing the grayish white ampule Return the protective ampule sleeve to the reaction tube with the gra...

Page 10: ...present depress the plunger again and draw a second sample slowly Remove the syringe from the oil sample and wipe any excess oil from the outside of the syringe with a tissue wipe be sure not to blot...

Page 11: ...bubble and perhaps foam during this time DO NOT BREAK THE GREEN AMPULE NOTE If the test sample is very thick and or viscous shake the reaction tube again for 30 seconds and let stand for an additiona...

Page 12: ...TUBE WHILE INSERTING THE BLACK SEPTUM DOING THIS WILL RESULT IN LOW READINGS 4 REACTION Break the bottom grayish white ampule in the reaction tube by compressing the sides of the tube Mix thoroughly...

Page 13: ...3 REACTION Break the bottom gray ampule in the tube by compressing the sides of the tube Mix thoroughly by shaking the tube vigorously for 30 seconds The mixture will be seen to bubble and perhaps fo...

Page 14: ...ight The difference in the two weights becomes the sample weight S2 to be used in equation 2 below Insert the black septum hole on top into the reaction tube by gripping the tube firmly and pressing d...

Page 15: ...0 0 50 01 50 61 1 HydroSCOUT results converted to w using sample density Analysis of Used Oil Certified Reference Materials on Programs A and B2 CRM Certified Value wt HydroSCOUT1 wt Karl Fischer wt E...

Page 16: ...Oil Light 26 53 08 236 202 007 448 410 043 838 778 026 Shell Turbo T32 23 34 019 220 223 04 433 403 01 825 800 038 Ideal Turbine Oil AW 32 39 43 072 210 208 013 400 435 005 759 842 013 Renolin Turbin...

Page 17: ...0 0 2 051 87 101 038 175 166 057 349 374 009 696 712 027 1397 1528 015 Valvoline Non Detergent SAE 30 0 30 059 88 96 042 175 136 061 350 368 028 701 784 02 1401 1497 041 Oilzum Non Detergent SAE 50 0...

Page 18: ...172 126 047 343 289 035 687 602 039 1374 1313 015 Quaker State SAE 30 0 7 018 88 105 06 176 183 104 352 333 044 703 597 04 1406 1491 018 Quaker State Proline 10W 40 Used 0 26 079 89 195 076 178 294 0...

Page 19: ...0 18 68 0 0184 29 9 26 63 0 0081 Topsoil 0 0 21 5 0 5 13 0 0377 10 1 10 33 0 0248 20 0 19 50 0 0026 Sand 0 0 22 5 1 5 30 0 0266 10 1 10 30 0 0274 15 0 15 42 0 0304 Clay Loam I 0 0 5 0 4 47 0 0354 10 2...

Page 20: ...luid D Phenolic resin containing approximately 10 unreacted Phenol Fluid E Pigmented polyester polyol resin system containing MEK Toluene Ethyl Acetate PM Acetate and Ethyl 3 Ethoxy Propionate Fluid F...

Page 21: ...wt G 0 15 wt wt 100 wt wt Using a 10 gram sample Using a 1 gram sample Program Storage Non Volatile Display 4 Digit 0 35 inch LCD with Program Indicator Battery One 9V Lithium included Use only Alkali...

Reviews: