Dexsil HydroSCOUT User Manual Download Page 17

14 

 

Analysis of Water Content in Various Fluids Using Program D 

 

Fluid Type

 

Karl Fischer (µg/mL)

 

HydroSCOUT (µg/mL)

 

C.V.

 

NAPA Non-Detergent SAE 30

 

0

 

17

 

.073

 

 

88

 

117

 

.069

 

 

176

 

217

 

.08

 

 

351

 

303

 

.071

 

 

702

 

661

 

.064

 

 

1405

 

1450

 

.048

 

Amalie Non-Detergent SAE 30

 

0

 

2

 

.051

 

 

87

 

101

 

.038

 

 

175

 

166

 

.057

 

 

349

 

374

 

.009

 

 

696

 

712

 

.027

 

 

1397

 

1528

 

.015

 

Valvoline Non-Detergent SAE 30

 

0

 

30

 

.059

 

 

88

 

96

 

.042

 

 

175

 

136

 

.061

 

 

350

 

368

 

.028

 

 

701

 

784

 

.02

 

 

1401

 

1497

 

.041

 

Oilzum Non-Detergent SAE 50

 

0

 

76

 

.051

 

 

89

 

126

 

.024

 

 

177

 

212

 

.06

 

 

355

 

404

 

.02

 

 

707

 

799

 

.009

 

 

1419

 

1595

 

.016

 

Gunk Heavy-Duty Brake Fluid

 

3835

 

4123

 

.01

 

 

4318

 

4323

 

.013

 

 

4586

 

4795

 

.009

 

 

5136

 

5341

 

.002

 

Shell Diala A

 

0

 

70

 

.039

 

 

88

 

121

 

.054

 

 

177

 

189

 

.055

 

 

353

 

372

 

.012

 

 

707

 

715

 

.041

 

 

1413

 

1506

 

.018

 

 

Summary of Contents for HydroSCOUT

Page 1: ...aims any implied warranties of merchantability or fitness for any particular purpose Further Dexsil Corporation reserves the right to revise this publication and to make changes from time to time in t...

Page 2: ...BROKEN GLASS AS IT CAN PENETRATE THE PLASTIC TUBE AND CUT FINGERS THE GRAY AMPULE IN THE TUBES CONTAINS CALCIUM HYDRIDE WHICH IS A FLAMMABLE SOLID AND IS WATER REACTIVE IN CASE OF ACCIDENTAL BREAKAGE...

Page 3: ...gram A Procedure for Testing Water Content up to 20 6 7 Program B Procedure for Testing Water Content up to 100 7 Program C D E Procedure for Low Levels of Water 5 mL Sample Size 8 Program F Procedure...

Page 4: ...ogram is described below Table 1 HydroSCOUT Analysis Program Information Program Sample Matrix Types Tested1 Units MDL3 Maximum Reagents Needed Sample Size A Used Oil Liquid Used Oil Chemtool 250 Pain...

Page 5: ...the pressure function is a linear parameter the meter response value can be corrected to produce accurate results on all oil types The HydroSCOUT response on low water level oils has been documented...

Page 6: ...thod D 7358 In order to achieve a result based on the weight of the sample the sample must be weighed into the reaction vessel Not only does this allow for a weight based result but it improves accura...

Page 7: ...to G followed by OFF and then back to A 0 20 again Press the READ key at any point to fix the selection at the displayed program or to turn the meter off when OFF is displayed To begin using the selec...

Page 8: ...i e programs C D or E ethylene glycol and acids have been found to produce a positive result Ethylene glycol responds at approximately 7 of the volumetric equivalent Error Code Explanation Likely Caus...

Page 9: ...ule from the cardboard carton and insert it fully into the top of the protective ampule sleeve containing the grayish white ampule Return the protective ampule sleeve to the reaction tube with the gra...

Page 10: ...present depress the plunger again and draw a second sample slowly Remove the syringe from the oil sample and wipe any excess oil from the outside of the syringe with a tissue wipe be sure not to blot...

Page 11: ...bubble and perhaps foam during this time DO NOT BREAK THE GREEN AMPULE NOTE If the test sample is very thick and or viscous shake the reaction tube again for 30 seconds and let stand for an additiona...

Page 12: ...TUBE WHILE INSERTING THE BLACK SEPTUM DOING THIS WILL RESULT IN LOW READINGS 4 REACTION Break the bottom grayish white ampule in the reaction tube by compressing the sides of the tube Mix thoroughly...

Page 13: ...3 REACTION Break the bottom gray ampule in the tube by compressing the sides of the tube Mix thoroughly by shaking the tube vigorously for 30 seconds The mixture will be seen to bubble and perhaps fo...

Page 14: ...ight The difference in the two weights becomes the sample weight S2 to be used in equation 2 below Insert the black septum hole on top into the reaction tube by gripping the tube firmly and pressing d...

Page 15: ...0 0 50 01 50 61 1 HydroSCOUT results converted to w using sample density Analysis of Used Oil Certified Reference Materials on Programs A and B2 CRM Certified Value wt HydroSCOUT1 wt Karl Fischer wt E...

Page 16: ...Oil Light 26 53 08 236 202 007 448 410 043 838 778 026 Shell Turbo T32 23 34 019 220 223 04 433 403 01 825 800 038 Ideal Turbine Oil AW 32 39 43 072 210 208 013 400 435 005 759 842 013 Renolin Turbin...

Page 17: ...0 0 2 051 87 101 038 175 166 057 349 374 009 696 712 027 1397 1528 015 Valvoline Non Detergent SAE 30 0 30 059 88 96 042 175 136 061 350 368 028 701 784 02 1401 1497 041 Oilzum Non Detergent SAE 50 0...

Page 18: ...172 126 047 343 289 035 687 602 039 1374 1313 015 Quaker State SAE 30 0 7 018 88 105 06 176 183 104 352 333 044 703 597 04 1406 1491 018 Quaker State Proline 10W 40 Used 0 26 079 89 195 076 178 294 0...

Page 19: ...0 18 68 0 0184 29 9 26 63 0 0081 Topsoil 0 0 21 5 0 5 13 0 0377 10 1 10 33 0 0248 20 0 19 50 0 0026 Sand 0 0 22 5 1 5 30 0 0266 10 1 10 30 0 0274 15 0 15 42 0 0304 Clay Loam I 0 0 5 0 4 47 0 0354 10 2...

Page 20: ...luid D Phenolic resin containing approximately 10 unreacted Phenol Fluid E Pigmented polyester polyol resin system containing MEK Toluene Ethyl Acetate PM Acetate and Ethyl 3 Ethoxy Propionate Fluid F...

Page 21: ...wt G 0 15 wt wt 100 wt wt Using a 10 gram sample Using a 1 gram sample Program Storage Non Volatile Display 4 Digit 0 35 inch LCD with Program Indicator Battery One 9V Lithium included Use only Alkali...

Reviews: