Dexsil HydroSCOUT User Manual Download Page 14

11 

 

Procedure G – Procedure for the Analysis of Paints and Coatings

  (cont’d) 

(Reagent System: HS-HPT)

 

Procedure for Expected Water Content Greater than 15% (with Dilution)

 

 

1.

 

PREPARATION:

  Remove a dilution vial, two sampling syringes, two caps and a white-capped tube from the 

box.  Check the contents to ensure that all items are present and intact.  Remove the white cap from the reaction 
tube  and  discard;  remove  the  cap  from  the  dilution  vial.    Insert  one  green  water  ampule  into  the  top  of  the 
polypropylene  ampule  sleeve  and  replace  the  ampule  sleeve  into  the  reaction  tube  with  the  grayish-white 
ampule on the bottom. 

2.

 

SAMPLE  INTRODUCTION:

  Remove  the  cap  from  an  empty  sampling  syringe.    Work  the  plunger  on  the 

empty  sampling  syringe  a  few  times  to  ensure  that  it  slides  easily.    Place  the  tip  of  the  syringe  into  the  paint 
sample to be tested and slowly pull back on the plunger until it reaches the stop and cannot be pulled further.  
When  drawing  up  the  sample,  make  sure  there  are  no  air  bubbles  in  the  syringe.    If  air  bubbles  are  present, 
depress the plunger again and draw a  second  sample  slowly.

  

Remove  the  syringe  from the paint  sample and 

wipe  any  excess  paint  from  the  outside  of  the  syringe  with  a  tissue  wipe,  be  sure  not  to  blot  the  tip  of  the 
syringe.  Place a syringe cap on the syringe and weigh; record the weight (gross sample weight) to the nearest 
0.1 mg. 

3.

 

SAMPLE  DILUTION:

  Remove  the  syringe  cap  and  insert  the  sampling  syringe  into  the  dilution  vial  and 

dispense the paint sample by depressing the plunger.  (Do not allow the syringe to contact the dilution solvent 
when dispensing the sample.)  Replace the syringe cap and reweigh the syringe and record the weight (sample 
tare weight).  The difference in the two weights becomes the sample weight S

1

 to be used in equation 2 below.  

Replace the  white cap on the dilution  vial and shake the vial vigorously for at least 10 seconds.  Remove the 
vial  cap  and  quickly,  using  a  new  sampling  syringe,  draw  up  a  sample  from  the  solution.    Wipe  any  excess 
solution from the outside of the syringe, cap with a fresh cap and weigh the syringe.  Record this weight (diluted 
sample  gross  weight).    Dispense  this  sample  into  the  reaction  tube.    Replace  the  cap  and  reweigh  the  syringe 
and  record  this  weight  (diluted  sample  tare  weight).    The  difference  in  the  two  weights  becomes  the  sample 
weight  S

2

  to  be  used  in  equation  2  below.  Insert  the  black  septum  (hole  on  top)  into  the  reaction  tube  by 

gripping  the  tube  firmly  and  pressing  down  on  the  septum  with  your  thumb  until  flush  with  top  of  the  tube.  

TIGHTLY

 cap the tube with the holed, white cap

.  NOTE: DO NOT SQUEEZE THE SIDES OF THE TUBE 

WHILE INSERTING THE BLACK SEPTUM.  DOING THIS WILL RESULT IN LOW READINGS.

4.

 

REACTION:

    Break  the  bottom  (gray)  ampule  in  the  tube  by  compressing  the  sides  of  the  tube.    Mix 

thoroughly  by  shaking  the  tube  vigorously  for  30  seconds.    The  mixture  will  be  seen  to  bubble  and  perhaps 
foam.  Allow the reaction to proceed for 2 minutes.  Shake the reaction tube again for 10 seconds and let stand 
for an additional 1 minute. 

5.

 

ANALYSIS:

 Press the <ON> key to activate the meter.  Ensure the meter is set for "Program G" (See “Selecting 

a Program” in the  manual on how to change programs).  Once the  meter displays “tUbE”, insert the reaction 
tube firmly into the opening at the bottom of the  meter.  The hole in the  white cap  must be lined up  with the 
metal pin extending from the inside of the hole in the meter.  Push the reaction tube all the way into the meter.  
Press the <READ> key on the face of the meter.  The display will flash the word "CALC," then the results will 
displayed in 

milligrams

 of water in the diluted sample.  Record the results.  

KEEP THE METER AND TUBE 

UPRIGHT AT ALL TIMES.

 

6.

 

DISPOSAL:

    Carefully  remove  the  reaction  tube  from  the  meter.    Vent  tube  by  inserting  completely  into 

venting  cap  and  remove  slowly.   

REACTION  TUBE  AND  CAP  MUST  BE  UPRIGHT  AND  POINTED 

AWAY FROM USER AND BYSTANDERS

.  Tap the tube on the bench top to shake any loose glass to the 

bottom of the tube, then break the top (green) ampule in the reaction tube by compressing the sides of the tube.  
Shake the tube for 15 seconds.  After 5 minutes, vent the tube again.  Discard the tube and vial as laboratory 
waste.  

DO NOT REMOVE WHITE CAP.

 

7.

 

CALCULATION:

  The meter is programmed to display the weight, in 

milligrams

, of the water contained in the 

sample  introduced  into  the  reaction  tube  (In  this  case  the  diluted  sample).    Calculate  the  water  content  in  the 
original paint sample in weight percent using the following equation: 

 

Water in paint, wt. percent =   (R/S

2

)(1+(2.15/S

1

))/10 

 

 

 

 

(2) 

where: 

R = Meter reading in 

milligrams

 of water 

S

1

 = actual weight of whole paint sample, in 

grams

 

S

2

 = actual weight of diluted paint sample, in 

grams

 

2.15 = Weight of the dilution solvent, in 

grams

 

Summary of Contents for HydroSCOUT

Page 1: ...aims any implied warranties of merchantability or fitness for any particular purpose Further Dexsil Corporation reserves the right to revise this publication and to make changes from time to time in t...

Page 2: ...BROKEN GLASS AS IT CAN PENETRATE THE PLASTIC TUBE AND CUT FINGERS THE GRAY AMPULE IN THE TUBES CONTAINS CALCIUM HYDRIDE WHICH IS A FLAMMABLE SOLID AND IS WATER REACTIVE IN CASE OF ACCIDENTAL BREAKAGE...

Page 3: ...gram A Procedure for Testing Water Content up to 20 6 7 Program B Procedure for Testing Water Content up to 100 7 Program C D E Procedure for Low Levels of Water 5 mL Sample Size 8 Program F Procedure...

Page 4: ...ogram is described below Table 1 HydroSCOUT Analysis Program Information Program Sample Matrix Types Tested1 Units MDL3 Maximum Reagents Needed Sample Size A Used Oil Liquid Used Oil Chemtool 250 Pain...

Page 5: ...the pressure function is a linear parameter the meter response value can be corrected to produce accurate results on all oil types The HydroSCOUT response on low water level oils has been documented...

Page 6: ...thod D 7358 In order to achieve a result based on the weight of the sample the sample must be weighed into the reaction vessel Not only does this allow for a weight based result but it improves accura...

Page 7: ...to G followed by OFF and then back to A 0 20 again Press the READ key at any point to fix the selection at the displayed program or to turn the meter off when OFF is displayed To begin using the selec...

Page 8: ...i e programs C D or E ethylene glycol and acids have been found to produce a positive result Ethylene glycol responds at approximately 7 of the volumetric equivalent Error Code Explanation Likely Caus...

Page 9: ...ule from the cardboard carton and insert it fully into the top of the protective ampule sleeve containing the grayish white ampule Return the protective ampule sleeve to the reaction tube with the gra...

Page 10: ...present depress the plunger again and draw a second sample slowly Remove the syringe from the oil sample and wipe any excess oil from the outside of the syringe with a tissue wipe be sure not to blot...

Page 11: ...bubble and perhaps foam during this time DO NOT BREAK THE GREEN AMPULE NOTE If the test sample is very thick and or viscous shake the reaction tube again for 30 seconds and let stand for an additiona...

Page 12: ...TUBE WHILE INSERTING THE BLACK SEPTUM DOING THIS WILL RESULT IN LOW READINGS 4 REACTION Break the bottom grayish white ampule in the reaction tube by compressing the sides of the tube Mix thoroughly...

Page 13: ...3 REACTION Break the bottom gray ampule in the tube by compressing the sides of the tube Mix thoroughly by shaking the tube vigorously for 30 seconds The mixture will be seen to bubble and perhaps fo...

Page 14: ...ight The difference in the two weights becomes the sample weight S2 to be used in equation 2 below Insert the black septum hole on top into the reaction tube by gripping the tube firmly and pressing d...

Page 15: ...0 0 50 01 50 61 1 HydroSCOUT results converted to w using sample density Analysis of Used Oil Certified Reference Materials on Programs A and B2 CRM Certified Value wt HydroSCOUT1 wt Karl Fischer wt E...

Page 16: ...Oil Light 26 53 08 236 202 007 448 410 043 838 778 026 Shell Turbo T32 23 34 019 220 223 04 433 403 01 825 800 038 Ideal Turbine Oil AW 32 39 43 072 210 208 013 400 435 005 759 842 013 Renolin Turbin...

Page 17: ...0 0 2 051 87 101 038 175 166 057 349 374 009 696 712 027 1397 1528 015 Valvoline Non Detergent SAE 30 0 30 059 88 96 042 175 136 061 350 368 028 701 784 02 1401 1497 041 Oilzum Non Detergent SAE 50 0...

Page 18: ...172 126 047 343 289 035 687 602 039 1374 1313 015 Quaker State SAE 30 0 7 018 88 105 06 176 183 104 352 333 044 703 597 04 1406 1491 018 Quaker State Proline 10W 40 Used 0 26 079 89 195 076 178 294 0...

Page 19: ...0 18 68 0 0184 29 9 26 63 0 0081 Topsoil 0 0 21 5 0 5 13 0 0377 10 1 10 33 0 0248 20 0 19 50 0 0026 Sand 0 0 22 5 1 5 30 0 0266 10 1 10 30 0 0274 15 0 15 42 0 0304 Clay Loam I 0 0 5 0 4 47 0 0354 10 2...

Page 20: ...luid D Phenolic resin containing approximately 10 unreacted Phenol Fluid E Pigmented polyester polyol resin system containing MEK Toluene Ethyl Acetate PM Acetate and Ethyl 3 Ethoxy Propionate Fluid F...

Page 21: ...wt G 0 15 wt wt 100 wt wt Using a 10 gram sample Using a 1 gram sample Program Storage Non Volatile Display 4 Digit 0 35 inch LCD with Program Indicator Battery One 9V Lithium included Use only Alkali...

Reviews: