USER Manual
30
Figure 7.7 X-RAY generating tube exposure ready status display
If exposure ready is complete, press READY button on bottom part of right side
and change to X-RAY exposure standby status as shown in Figure 7.8.
If READY button is pressed, the button turns into red CANCEL button, and it will
be standing by until X-RAY output is detected. Standby time can be set up within
5 ~ 180 seconds range, and setting can be done on SETUP panel (Refer to 7.6.3).
If READY button is pressed and turned into X-RAY exposure standby status, panel
navigation and setting change related buttons, except CANCEL button, are
deactivated.
After locating flat face of intra oral sensor to the closest distance inside
radiograph target teeth, position radiation field of ADX4000W parallel with
sensor face, and to the closest distance possible.
If aiming is completely, press EX button to activate exposure. Ex button should
not be released until exposure is complete, and if button is released in the
middle of exposure, exposure is suspended immediately. Exposure is started with
a beeping sound, and EX button on front part of main body changes to yellow
color.
Exposure ready is not complete
Exposure ready is complete.