Inspection items before repair request
9.1
Confirm items before requesting inspection
If abnormality is found in product, confirm following items before inspection request.
Symptom
Steps to take
Po
w
er
d
ef
ec
t
After turning on power,
both sound or LCD
screen are not turned on
Confirm battery is mounted correctly and replace
with charged battery.
If not using product for a long time, separate
battery from product to prevent battery discharging
LCD screen is not
displaying after power
on sound is heard
While system is booted, LCD screen may not be
displayed for a while. Wait until product UI is
displayed by PC booting completion after passing
DEXCOWIN logo. If screen is not displayed after a
few minutes, contact designated A/S center.
While using (including
X-RAY exposure), if
power turns off
automatically
Check battery residual quantity.
This product supports automatic power turn off
function to reduce battery consumption.
Exposure defect
When X-RAY is working
for exposure, and
exposure sound does
not go off
Be sure to set exposure setting value before X-RAY
exposure, and press READY button for exposure.
Confirm battery residual quantity, and replace with
charged battery. X-RAY exposure may not be
performed if battery level is low.
Check if PC is connected through USB cable
normally and separate the USB cable.
Check if the panel of LCD display is X-RAY mode. X-
RAY exposures only with X-RAY panel.