Dexcom G4 PLATINUM Professional System User’s Guide
|
101
Status Symbols
(continued from page before)
Glucose
Reading
Error
This symbol means the receiver does not understand the sensor signal
but is likely to recover. This symbol is related to the sensor only. You
should wait for more prompts and
do not enter
any blood glucose
values when you see this symbol (see Chapter 13, Section 13.4, System
Glucose Error, for more information).
Out of
Range
This symbol means the receiver and sensor/transmitter are not
communicating. Make sure the receiver and sensor/transmitter are
within 20 feet of each other without obstruction (see Chapter 1, Section
1.4, Transmitter Overview).
Wait 15
Minutes
Calibration
Error
This symbol means the sensor cannot calibrate right now. If you see
this screen, enter at least one more calibration blood glucose value after
about 10-15 minutes. If the sensor still cannot calibrate after that, the
sensor needs to be removed and a new sensor needs to be inserted.
Wait 1
Hour
Calibration
Error
This symbol means the sensor is not calibrating correctly. If you see
this screen, wait a minimum of one hour and then enter one more blood
glucose value. If no readings display on the receiver after this, the
sensor needs to be removed and a new sensor needs to be inserted.
Wait
This symbol means the receiver has detected a potential significant
problem with the sensor signal and may result in a sensor failure. You
should wait about 30 minutes for more prompts. Do not enter any blood
glucose values when you see this symbol.
Status symbols show in the upper right of the trend graph.
Summary of Contents for G4 Platinum Professional
Page 13: ...12 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1...
Page 25: ...24 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 2...
Page 36: ...Dexcom G4 PLATINUM Professional System User s Guide 35 3 chapter three RISKS AND BENEFITS...
Page 37: ...36 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 1 3...
Page 43: ...42 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 4...
Page 51: ...50 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 5...
Page 65: ...64 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 6...
Page 81: ...80 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 7...
Page 93: ...92 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10 8...
Page 104: ...Dexcom G4 PLATINUM Professional System User s Guide 103 9 chapter nine ALERTS ALARM PROFILES...
Page 105: ...104 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 9...
Page 127: ...126 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 10...
Page 141: ...140 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 11...
Page 149: ...148 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 12...
Page 164: ...Dexcom G4 PLATINUM Professional System User s Guide 163 13 chapter thirteen TROUBLESHOOTING...
Page 165: ...164 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 13...
Page 183: ...182 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 14...
Page 218: ...Dexcom G4 PLATINUM Professional System User s Guide 217 15 chapter fifteen USER ASSISTANCE...
Page 219: ...218 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 15...
Page 222: ...Dexcom G4 PLATINUM Professional System User s Guide 221 16 chapter sixteen WARRANTY...
Page 223: ...222 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 16...
Page 233: ...232 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 17...
Page 236: ...Dexcom G4 PLATINUM Professional System User s Guide 235 18 chapter eighteen APPENDIX...
Page 237: ...236 Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System 18...