IMPORTANT!

 Do not provide these setup 

instructions to the patient. Only qualified 

medical professionals should access these 

device settings.

Follow these instructions to access the device 

settings and prepare the device for patient use. 

All settings have pre-set values as shown below.

Other Settings

Filter Cleaning Reminder

 – Displays a reminder to the patient to clean the filter 

every 50 hours of use, or approximately every 10 days. This feature is controlled by 

the notifications setting and is preset to enabled. 

Other setup information

•  This device features automatic altitude adjustment for altitudes up to 2743m 

(9000ft) and needs no manual adjustment.

•  This device features a pressure sensor therefore no fine pressure adjustments are 

needed.

•  The SmartLink Data Management System may be used for additional reminders and 

notifications.

•  Clear compliance meter by pressing Down key for eight seconds with compliance 

meter shown (value will go to zero).

All device settings are immediately updated in the device as you make adjustments. 

When finished press the ON/OFF button once to exit the Clinical Setup Menu. 

Second

 -

 While holding the 

two buttons, press the 

ON/

OFF

 button. The display will 

show “Clinical Menu”.  You 

can now access the Clinical 

Setup Menu.

Use the 

LEFT

 but-

ton and 

RIGHT

 

button to navigate 

through the menu.

Use the 

UP

 button 

and 

DOWN

 button to 

adjust the settings.

First -

 With the device pow-

ered and turned off, press and 

hold the 

DOWN

 and 

DELAY

 

buttons. 

DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881

Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688

 

EN

Menu Item

Description

Factory Default

AutoAdjust Mode Clinical Menu

Language

Selection

Select device language. 

 

English

Operating 

Mode

Adjust mode to AutoAdjust or CPAP as prescribed by 

doctor.  

Range: AutoAdjust, CPAP

DV54: AutoAdjust

DV53: CPAP

Upper 

Limit

Maximum pressure level that will be administered by the 

device.

Range: Lower Limit to 20 cmH

2

0, increments of 0.5

15.0 cmH

2

0

Lower 

Limit

Minimum pressure level that will be administered by the 

device.

Range: 3.0 cmH

2

0 to Upper Limit, increments of 0.5

5.0 cmH

2

0

Delay 

Time

During delay the pressure remains at Lower Limit and 

the algorithm continues to detect events, but no pres-

sure changes are made.

Range: 0-45 minutes, increments of 5

20 min

Pressure 

Unit

The device operating pressure may be displayed in one 

of three pressure units: cmH

2

O, hPa or mbar. 

cmH

2

0

Tubing 

Length

The device is supplied with air supply tubing of 1.8 m 

 

(6 ft) in length. The device may also be operated with air 

supply tubing of 3 m (10 ft). This setting should be 

updated to match the tubing length for maximum 

pressure accuracy. 

1.8 m (6 ft)

Compliance 

Meter

Cumulative hours of “while breathing” time.

N/A

Advanced 

Menu

Press the 

 button to enter. Reference the Advanced 

Submenu section for details.

N/A

Enable 

Menu

Press the 

 button to enter. Reference the Enable 

Submenu section for details.

N/A

Lockout 

Menu

Press the 

 button to enter. Reference the Lockout 

Submenu section for details.

N/A

CPAP Mode Clinical Menu

Language

Selection

Select device language. 

 

English

Operating 

Mode

Adjust mode to AutoAdjust or CPAP as prescribed by 

doctor.  

Range: AutoAdjust, CPAP

DV54: AutoAdjust

DV53: CPAP

Set 

Pressure

Set to CPAP pressure prescribed by doctor. 

Range: 3-20 cmH

2

O

10.0 cmH

2

0

Delay 

Time

Select period of time to stay at delay pressure. This 

delay is patient activated.

Range: 0-45 minutes, increments of 5

20 min

Delay 

Pressure

If prescribed, set to a level that is comfortable for the 

patient to fall asleep. This level of pressure is maintained 

for the first portion of the delay time and rises to the set 

pressure in the last 10 minutes of the delay time.

Range:  3 cmH

2

O to CPAP, increments of 0.5

4.0 cmH

2

0

Pressure 

Unit

The device operating pressure may be displayed in one 

of three pressure units: cmH

2

O, hPa or mbar. 

cmH

2

0

Menu Item

Description

Factory Default

Tubing 

Length

The device is supplied with air supply tubing of 1.8 m 

 

(6 ft) in length. The device may also be operated with air 

supply tubing of 3 m (10 ft). This setting should be 

updated to match the tubing length for maximum 

pressure accuracy. 

1.8 m (6 ft)

Compliance 

Meter

Cumulative hours of “while breathing” time.

N/A

Advanced 

Menu

Press the 

 button to enter. Reference the Advanced 

Submenu section for details.

N/A

Enable 

Menu

Press the 

 button to enter. Reference the Enable 

Submenu section for details.

N/A

Lockout

Menu

Press the 

 button to enter. Reference the Lockout 

Submenu section for details.

N/A

Advanced Submenu

Apnea %

The absolute value below which the respiratory flow 

must fall to be considered as an apnea.  

Range:  0-20%, increments of 5

10%

Apnea 

Duration

The number of seconds the respiratory flow must 

remain below the apnea percent to be detected as an 

apnea.  

Range:  6-150 seconds, increments of 2

10 seconds

Hypopnea % The percent which the respiratory flow must fall from its 

normal value to be considered as a hypopnea.  

Range:  30%-70%, increments of 5

50%

Hypopnea 

Duration

The number of seconds the respiratory flow must 

remain below the hypopnea percent to be detected as a 

hypopnea. Range:  6-150 seconds, increments of 2

10 seconds

Exit 

Submenu

Press the 

 button to exit and return to Clinical Menu.

N/A

Enable Submenu

Auto-ON

This option controls the Auto-ON feature, which 

automatically starts the flow of air after breathing once 

or twice into the mask

Enabled

Auto-OFF

This option controls the Auto-OFF feature, which 

automatically stops the flow of air after taking off the 

mask. 

Disabled

Mask Fit Chk Displays a reminder to the patient to check the fit of the 

mask if airflow was greater than 95L/Min for more than 

10% of the previous therapy session (including vent 

leak). The previous therapy session must be greater than 

3 hours or a reminder will not be displayed.

Enabled

Notifications

Setting to enable mask/filter check notification.

Enabled

Exit 

Submenu

Press the 

 button to exit and return to Clinical Menu.

N/A

Lockout Submenu

Delay Time 

Lockout

Used to allow patient access to CPAP delay time setting.

Range:  Locked, Unlocked

Unlocked

Tube Length 

Lockout

Used to allow patient access to tubing length setting.

Range:  Locked, Unlocked

Unlocked

Exit 

Submenu

Press the 

 button to exit and return to Clinical Menu.

N/A   

©2007 DeVilbiss Healthcare 12.07

DV54D-125 Rev. A

Quick Set-Up Guide 

IntelliPAP™AutoAdjust • SleepCube™ AutoAdjust  • SleepCube Plus™ CPAP

Summary of Contents for IntelliPAP AutoAdjust

Page 1: ...enu Language Selection Select device language English Operating Mode Adjust mode to AutoAdjust or CPAP as prescribed by doctor Range AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP Set Pressure Set to CPAP...

Page 2: ...lecci n de Idioma Seleccionar el idioma del dispositivo Ingl s Modo Operativo Ajuste el modo en AutoAdjust Auto ajuste o CPAP seg n lo indicado por el m dico Rango AutoAdjust CPAP DV54 AutoAdjust DV53...

Page 3: ...enu clinique Mode CPAP S lection de la langue S lection de la langue d affichage Anglais Mode de fonc tionnement R glage sur AutoAdjust ou sur CPAP selon les instructions du m decin Plage AutoAdjust C...

Page 4: ...djust oder CPAP je nach Anordnung des Arztes Optionen AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP Druckeinheit CPAP Druck auf den vom Arzt verschriebenen Wert einstel len Optionen 3 20 cmH2O 10 0 cmH20...

Page 5: ...CPAP Menu clinico Selezione della lingua Seleziona la lingua del dispositivo Inglese Modal operativa Regola la modalit su Auto Adjust o CPAP come prescritto dal medico Gamma AutoAdjust CPAP DV54 AutoA...

Page 6: ...t Engels Bedienings modus Stel de modus in op AutoAdjust of CPAP zoals voorge schreven door de arts Bereik AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP Ingestelde druk Stel de CPAP druk in die is voorges...

Page 7: ...ll in l ge Autojust eller CPAP enligt l kares ordina tion Omf ng Autojust CPAP DV54 AutoAdjust DV53 CPAP Inst llning av trycket St ll in det CPAP tryck som ordinerats av l kare Omf ng 3 20 cm H2O 10...

Page 8: ...r yksen mukaisesti Vaihteluv li AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP Paineasetus Aseta CPAP paineeseen l k rin m r yksen mukai sesti Vaihteluv li 3 20 cm H2O 10 0 cm H20 Viiveaika Valitse painee...

Page 9: ...l AutoAdjust eller CPAP if lge l gens ordinering Omr de AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP Indstillet tryk Indstil p det CPAP tryk der er ordineret af din l ge Omr de 3 20 cmH2O 10 0 cmH20 Fors...

Page 10: ...yrsspr k Norsk Driftsmodus Juster modus til AutoAdjust eller CPAP som foreskrevet av lege Verdiomr de AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP Innstille trykk Innstill til CPAP trykk foreskrevet av l...

Page 11: ...CPAP AutoAdjust CPAP DV54 AutoAdjust DV53 CPAP 20 cmH20 0 5 15 0 cmH20 3 0 cmH20 0 5 5 0 cmH20 0 45 5 20 cmH2O hPa mbar cmH20 1 8 6 3 10 1 8 6 CPAP AutoAdjust CPAP AutoAdjust CPAP DV54 AutoAdjust DV5...

Reviews: