IMPORTANT!
Do not provide these setup
instructions to the patient. Only qualified
medical professionals should access these
device settings.
Follow these instructions to access the device
settings and prepare the device for patient use.
All settings have pre-set values as shown below.
Other Settings
Filter Cleaning Reminder
– Displays a reminder to the patient to clean the filter
every 50 hours of use, or approximately every 10 days. This feature is controlled by
the notifications setting and is preset to enabled.
Other setup information
• This device features automatic altitude adjustment for altitudes up to 2743m
(9000ft) and needs no manual adjustment.
• This device features a pressure sensor therefore no fine pressure adjustments are
needed.
• The SmartLink Data Management System may be used for additional reminders and
notifications.
• Clear compliance meter by pressing Down key for eight seconds with compliance
meter shown (value will go to zero).
All device settings are immediately updated in the device as you make adjustments.
When finished press the ON/OFF button once to exit the Clinical Setup Menu.
Second
-
While holding the
two buttons, press the
ON/
OFF
button. The display will
show “Clinical Menu”. You
can now access the Clinical
Setup Menu.
Use the
LEFT
but-
ton and
RIGHT
button to navigate
through the menu.
Use the
UP
button
and
DOWN
button to
adjust the settings.
First -
With the device pow-
ered and turned off, press and
hold the
DOWN
and
DELAY
buttons.
DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881
Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688
EN
Menu Item
Description
Factory Default
AutoAdjust Mode Clinical Menu
Language
Selection
Select device language.
English
Operating
Mode
Adjust mode to AutoAdjust or CPAP as prescribed by
doctor.
Range: AutoAdjust, CPAP
DV54: AutoAdjust
DV53: CPAP
Upper
Limit
Maximum pressure level that will be administered by the
device.
Range: Lower Limit to 20 cmH
2
0, increments of 0.5
15.0 cmH
2
0
Lower
Limit
Minimum pressure level that will be administered by the
device.
Range: 3.0 cmH
2
0 to Upper Limit, increments of 0.5
5.0 cmH
2
0
Delay
Time
During delay the pressure remains at Lower Limit and
the algorithm continues to detect events, but no pres-
sure changes are made.
Range: 0-45 minutes, increments of 5
20 min
Pressure
Unit
The device operating pressure may be displayed in one
of three pressure units: cmH
2
O, hPa or mbar.
cmH
2
0
Tubing
Length
The device is supplied with air supply tubing of 1.8 m
(6 ft) in length. The device may also be operated with air
supply tubing of 3 m (10 ft). This setting should be
updated to match the tubing length for maximum
pressure accuracy.
1.8 m (6 ft)
Compliance
Meter
Cumulative hours of “while breathing” time.
N/A
Advanced
Menu
Press the
button to enter. Reference the Advanced
Submenu section for details.
N/A
Enable
Menu
Press the
button to enter. Reference the Enable
Submenu section for details.
N/A
Lockout
Menu
Press the
button to enter. Reference the Lockout
Submenu section for details.
N/A
CPAP Mode Clinical Menu
Language
Selection
Select device language.
English
Operating
Mode
Adjust mode to AutoAdjust or CPAP as prescribed by
doctor.
Range: AutoAdjust, CPAP
DV54: AutoAdjust
DV53: CPAP
Set
Pressure
Set to CPAP pressure prescribed by doctor.
Range: 3-20 cmH
2
O
10.0 cmH
2
0
Delay
Time
Select period of time to stay at delay pressure. This
delay is patient activated.
Range: 0-45 minutes, increments of 5
20 min
Delay
Pressure
If prescribed, set to a level that is comfortable for the
patient to fall asleep. This level of pressure is maintained
for the first portion of the delay time and rises to the set
pressure in the last 10 minutes of the delay time.
Range: 3 cmH
2
O to CPAP, increments of 0.5
4.0 cmH
2
0
Pressure
Unit
The device operating pressure may be displayed in one
of three pressure units: cmH
2
O, hPa or mbar.
cmH
2
0
Menu Item
Description
Factory Default
Tubing
Length
The device is supplied with air supply tubing of 1.8 m
(6 ft) in length. The device may also be operated with air
supply tubing of 3 m (10 ft). This setting should be
updated to match the tubing length for maximum
pressure accuracy.
1.8 m (6 ft)
Compliance
Meter
Cumulative hours of “while breathing” time.
N/A
Advanced
Menu
Press the
button to enter. Reference the Advanced
Submenu section for details.
N/A
Enable
Menu
Press the
button to enter. Reference the Enable
Submenu section for details.
N/A
Lockout
Menu
Press the
button to enter. Reference the Lockout
Submenu section for details.
N/A
Advanced Submenu
Apnea %
The absolute value below which the respiratory flow
must fall to be considered as an apnea.
Range: 0-20%, increments of 5
10%
Apnea
Duration
The number of seconds the respiratory flow must
remain below the apnea percent to be detected as an
apnea.
Range: 6-150 seconds, increments of 2
10 seconds
Hypopnea % The percent which the respiratory flow must fall from its
normal value to be considered as a hypopnea.
Range: 30%-70%, increments of 5
50%
Hypopnea
Duration
The number of seconds the respiratory flow must
remain below the hypopnea percent to be detected as a
hypopnea. Range: 6-150 seconds, increments of 2
10 seconds
Exit
Submenu
Press the
button to exit and return to Clinical Menu.
N/A
Enable Submenu
Auto-ON
This option controls the Auto-ON feature, which
automatically starts the flow of air after breathing once
or twice into the mask
Enabled
Auto-OFF
This option controls the Auto-OFF feature, which
automatically stops the flow of air after taking off the
mask.
Disabled
Mask Fit Chk Displays a reminder to the patient to check the fit of the
mask if airflow was greater than 95L/Min for more than
10% of the previous therapy session (including vent
leak). The previous therapy session must be greater than
3 hours or a reminder will not be displayed.
Enabled
Notifications
Setting to enable mask/filter check notification.
Enabled
Exit
Submenu
Press the
button to exit and return to Clinical Menu.
N/A
Lockout Submenu
Delay Time
Lockout
Used to allow patient access to CPAP delay time setting.
Range: Locked, Unlocked
Unlocked
Tube Length
Lockout
Used to allow patient access to tubing length setting.
Range: Locked, Unlocked
Unlocked
Exit
Submenu
Press the
button to exit and return to Clinical Menu.
N/A
©2007 DeVilbiss Healthcare 12.07
DV54D-125 Rev. A
Quick Set-Up Guide