EN - 7

A-6910

Caution– The DeVilbiss Reusable Nebulizer is dishwasher safe, but do not place any disposable nebulizer parts in an 

automatic dishwasher; doing so could cause damage.
Disinfect Daily:

1.  Using a clean container or bowl, soak items in three parts hot water (131ºF-149ºF/55ºC-65ºC) to one part white 

vinegar (>=5% acetic acid concentration) for 30 minutes (Fig. 7) OR use a medical bacterial-germicidal disinfectant 

available through your provider. Be sure to follow manufacturer’s instructions carefully.

2.  With clean hands, remove items from disinfectant solution, rinse under hot tap water, and air dry on a clean paper 

towel. Store in a Zip-lock® bag.

note– Do not towel dry nebulizer parts; this could cause contamination.

WaRNINg

To prevent possible risk of infection from contaminated cleaning solutions, always prepare fresh solution for each 

cleaning cycle and discard solution after each use.

3.  For reusable nebulizer only, clean daily in dishwasher (if available) using top shelf.

4.  Keep the outer surface of the tubing dust-free by wiping regularly. Nebulizer tubing does not have to be washed 

because only filtered air passes through it.

note– The Disposable Nebulizer will last at least 15 days and possibly longer, depending on usage. Proper cleaning will 

also help extend the life of the nebulizer. Because the nebulizer is disposable, we recommend that an extra nebulizer be 

kept on hand at all times. The DeVilbiss Reusable Nebulizer is dishwasher safe and may be cleaned and reused for up to 

6 months.

COmpRESSOR CLEaNINg

1.  With the unit off, unplug power cord from wall outlet.

2.  Wipe the outside of the compressor cabinet with a clean damp cloth every few days to keep dust-free.

DaNgER

Do not submerge in water; doing so will result in compressor damage.

FILTER ChaNgE

1.  Filter should be changed every six months or sooner if filter discolors.

2.  Remove filter (Fig. A) by pulling back the filter cover tab and then pulling out the filter from the unit. Discard old filter. 

The filter cover tab should remain attached to the device.

3.  Replace with a new filter. Additional filters should be purchased from your DeVilbiss provider.

4.  Push new filter into position, and replace filter cover.

Caution– Do not reuse or wash filter or substitute any other material such as cotton for a DeVilbiss air-inlet filter. It will 

result in compressor damage.

maINTENaNCE

All maintenance must be performed by a authorized DeVilbiss provider.

DaNgER

Electric shock hazard. Do not remove compressor cabinet. All disassembly and maintenance must be done by a 

authorized DeVilbiss provider.