If your finger is removed too early, the tester’s LEDs will 

 

blink three times to indicate that the test was not 

 

completed.  DO NOT touch any other metal while 

 

performing your test as this will affect your results.

NOTE: Often the initial intermittency will be failure of strain 

relief connection to resistor as simulated by ESD S1.1-1998 

paragraph 5.7 Bending Life Test.
4.   A “PASS” test result is indicated by illumination of 

 

the green LEDs.  A “FAIL LOW” test result is indicated 

 

by illumination of the red LEDs.  A “FAIL HIGH” test 

 

result is indicated by illumination of the yellow LEDs.

 

If your ESD test fails, check your wrist strap and 

 

footwear to ensure that they are functioning properly, 

 

being worn correctly and/or need to be replaced.

NOTE: Failures may be caused by dry skin or minimal sweat 

layer.  For wrist straps, try using an approved dissipative 

hand lotion such as 

Menda Reztore™ ESD Hand Lotion

 

to your wrist prior to use.  Footwear test results can be 

improved by taking a short walk to build a sweat layer for 

better conductivity.
The Combo Tester X3 may also be used to test smocks or 

garments that feature a grounding mechanism for operators 

using a coil cord connection.

Figure 9.  Bridging the test switch’s contacts to initiate 

the test

Calibration

The Combo Tester X3 is calibrated to standards traceable 

to NIST.  Frequency of recalibratrion should be based on 

the critical nature of those ESD sensitive items handled 

and the risk of failure for the ESD protective equipment and 

materials.  In general, we recommend that calibration be 

performed annually.
The accuracy of the Combo Tester X3 is specified as:
•  ±10% for < 1 Megohm range

•  ±10% for 1 Megohm and higher resistance ranges

•  ±20% for 1 Gigohm Range
A periodic check (once every 6 to 12 months) using a 

precision resistance box should be performed to verify proper 

operation.
The EMIT 

50424

 Limit Comparator is available for the 

convenient period testing of the Combo Tester X3.
The Limit Comparator allows the customer to perform NIST 

traceable calibration on the Combo Tester X3.  The Limit 

Comparator can be used on the floor shop within a few 

minutes virtually eliminating downtime, verifiying that the 

Combo Tester X3 is operating within tolerances.
See 

TB-6581

 for more information.

Using the 

19276

Combo Tester X3

Using the 

19277

Combo Tester X3 with Stand

Summary of Contents for 19276

Page 1: ...ods and equipment used for process monitoring and measurements The test equipment selected shall be capable of making the measurements defined in the Compliance Verification Plan Per ANSI ESD S1 1 section 6 1 3 Frequency of Functional Testing The wrist strap system should be tested daily to ensure proper electrical value Per ESD Handbook ESD TR20 20 section 5 3 2 4 2 Additional User Wrist Strap Te...

Page 2: ...r for the relay to activate I Buzzer Volume Adjustment Turn the trimpot clockwise to increase the buzzer volume and counter clockwise to decrease the volume J Ground Jack Insert the banana plug end of the included ground cord to this jack Connect the ring terminal end of the cord to equipment ground This connection will remove any static charge from the user before the test NOTE Failure to correct...

Page 3: ...4 Use the guide located on the bottom of the foot plate to route the cord out of the side This will prevent the foot plate cord from being accidentally tripped and unplugged 5 Insert the ground cord s banana plug into the ground jack located at the bottom of the tester Connect ground cord s ring terminal to equipment ground This connection will remove any static charge from the user before the tes...

Page 4: ... before the test NOTE Failure to correctly ground the Combo Tester X3 may result in damage not covered under warranty 7 Fit the dual foot plate into a position so that is flush with the front and top of the baseplate Secure the dual foot plate to the baseplate using the four included 4 pan head screws Figure 7 Securing the dual foot plate to the baseplate 8 Power the Combo Tester X3 using the incl...

Page 5: ...mproved by taking a short walk to build a sweat layer for better conductivity The Combo Tester X3 may also be used to test smocks or garments that feature a grounding mechanism for operators using a coil cord connection Figure 9 Bridging the test switch s contacts to initiate the test Calibration The Combo Tester X3 is calibrated to standards traceable to NIST Frequency of recalibratrion should be...

Page 6: ...ddress Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warranty will not apply to defects or damage due to accidents neglect misuse a...

Page 7: ...Mouser Electronics Authorized Distributor Click to View Pricing Inventory Delivery Lifecycle Information Desco 19276 19277 19255 ...

Reviews: