8 Endodontic Mode
Page 25
Change Torque-Stop Mode
:
15. Select the direction of the auto reverse option for endodontic procedures by
pressing the TORQUE-STOP MODE icon.
16. Continue to press the TORQUE-STOP MODE icon until.
The desired direction option is displayed. The choices will be:
Auto-Stop
Auto-Reverse
Auto-Reverse Forward
(See time value set up in the "Set Up Mode" page 15.)
Torque-Stop Mode allows the user to define the response of the system when the specified torque is reached
on the endodontic file.
Auto-Stop:
When the specified torque value is reached on the endodontic file, the handpiece will stop.
Auto-Reverse:
When the specified torque value is reached on the endodontic file, the handpiece will
reverse in direction until the user releases the foot pedal.
Auto-Reverse Forward
: When the specified torque value is reached on the endodontic file, the handpiece
will reverse in direction for the specified time and then return to normal forward rotation.
NOTE:
The torque value determines the point at which the direction of rotation will stop
in Auto-stop or reverse in direction when in Auto-reverse or Auto-reverse forward.
Storing Settings in Memory
:
17. Pressing the E1 button and holding it in until the signal sounds stores the options
set in menu option E1.
18. Follow the same procedure for setting E2, E3, and E4 if desired.