3
Quick Reference Guide
Diagnostic Display
ON/OFF
Illuminates when the Main Power
On/Off Power switch is in the “ON”
(I) position.
TURBO
Offers the ability to increase power
to the system by up to 25% with
the push of a button. Purple arrows
illuminate when in use.
BOOST
Illuminates when the Boost Mode
is activated by the Tap-On
TM
Wireless Foot Pedal. To activate,
fully depress Tap-On
TM
Foot Pedal
to the second position (all the way
to the floor). To deactivate, release
Tap-On
TM
Foot Pedal to the first
position.
PURGE BUTTON
Illuminates when Purge function
is activated. To activate Purge,
remove insert from handpiece, turn
the Handpiece Lavage Control to
maximum water flow, press the
Purge Button on the Diagnostic
Display. Water will purge through
system lines for two minutes.
To
deactivate during two minute cycle,
press Purge button again or press
Tap-On
TM
Foot Pedal.
SERVICE
Illuminates when the system is not
functioning properly. This display
has three distinct modes:
Slow blink (1 blink per
•
second) means the system is
not operating within factory
specifications.
Fast blink (3 blinks per second)
•
indicates an improper set-up.
Steady light indicates the system
•
is overheating.
LOW BATTERY
Illuminates when the Tap-On
TM
Foot Pedal battery power is
approaching end of life. Replace
batteries as instructed in
Section 7.9.
Power Control
Power Level Control
Turn knob to select ultrasonic power
level for operation. Turning the knob
clockwise increases the distance the
insert tip moves (the stroke) without
changing frequency; turning knob
counter-clockwise decreases the distance
the insert tip moves (the stroke) without
changing the frequency.
RINSE
Rinse mode is used during an ultrasonic
scaling procedure when lavage is required
to flush the procedural area. To activate,
turn Power Level Control Knob fully
counter-clockwise until a “click” is heard.
BLUE ZONE
Provides an extended low-power range
for improved patient comfort when
subgingivally scaling.