4 Operation
4.8 Position sensor with AimRight reusable sensor holder system
Dentsply Sirona
Operating Instructions Xios AE Sensor USB interface and sensors
56
67 29 433 D3780
D3780.201.01.01.02 09.2019
2.
Position the sensor directly on the dental arch.
Ä
The bite block of the sensor holder must be located between
the upper molars, parallel to the occlusal plane.
Ä
In order to avoid superimpositions, the sensor must be
positioned parallel to the dental arch line.
3.
Slide the localizer ring to the patient's face.
4.
Align the cone of the X-ray tube assembly parallel to the sensor
directly on the localizer ring.
5.
Release an X-ray exposure. Refer to sections Select the exposure
parameters for the X-ray tube assembly and Releasing the
exposure. Also observe the operating instructions for the X-ray tube
assembly.
After the exposure
1.
Ask the patient to open their mouth.
2.
Remove the sensor from the patient’s mouth.
3.
Remove the sensor from the hygienic protective sleeve. Follow the
instructions in the section Removing the hygienic protective sleeve
from the sensor. The hygienic protective sleeves must be disposed
of after the examination.
4.
Clean and sterilize the sensor holder and localizer ring.