7
Output
≥ 500 mW/cm
2
Shades A2; A3; A3,5; A4; B1; B3; C2; C3
10 sec
Shades O-A2; O-B3
20 sec
Curing time for 2 mm increments
4 Hygiene
4.1 Compules
®
Tips Gun
For reprocessing instructions please refer to the Instructions for Use of the Compules
®
Tips Gun,
which is available on our webpage at www.dentsply.eu/IFU. If requested, we will send you a free
printed copy of the Instructions for Use in the language you require within 7 days. You can use the
order form provided on our webpage for this purpose.
4.2 Compules
®
Tips – after intraoral use
4.3 Shade guide holder and individual tabs
3.6 Finishing and polishing
1. Contour the restoration using finishing burs or diamonds.
2. Use Enhance
®
finishing instruments and interproximal finishing strips. In patients with adequate
oral hygiene, the high final luster of the restoration comes with use.
For contouring, finishing, and/or polishing, follow the manufacturer’s Instructions for Use.
Cross-contamination.
Infection.
1. Do not reuse Compules
®
Tips after intraoral use and/or if contaminated.
2. Dispose the intraorally used and/or contaminated Compules
®
Tips in accordance
with local regulations.
Cross-contamination.
Infection.
1. The shade guide and the individual shade guide tabs cannot be reprocessed.
2. To prevent the shade guide and the individual shade guide tabs from exposure
to spatter or spray of body fluids or contaminated hands it is mandatory that the
shade guide and the individual shade guide tabs are handled with clean/disin-
fected gloves. Do not reuse shade guide and the individual shade guide tabs if
contaminated.
3. Dispose contaminated shade guide and the individual shade guide tabs in accord-
ance with local regulations.