6 Operation
6.2 Calibrating the unit
67 19 681 D3439
D3439.201.01.01.02 09.2019
39
Calibration procedure
1.
Select the symbol shown on the left in the touch display.
2.
Select
"Calibration" and then "Start".
3.
Register the calibration phantom on the machine (scanning the
code) and insert it in the machine.
4.
Close the door.
5.
Register the calibration pins (set 1) on the machine (RFID) and
insert them in the machine.
6.
Close the door.
Ä
The calibration for set 1 is performed. The automatic calibration
starts and lasts approx. 4 minutes. Wait until the calibration has
been completed.
7.
Remove the calibration pins from the machine.
8.
Close the door.
9.
Register the calibration pins (set 2) on the machine (RFID) and
insert them in the machine.
10. Close the door.
Ä
The calibration for set 2 is performed. The automatic calibration
starts and lasts approx. 4 minutes. Wait until the calibration has
been completed.
11. Remove the calibration pins from the machine.
12. Close the door.
13. Remove the calibration phantom.
14. Close the door.
Ä
The calibration is completed.