Dentsply Sirona
Operating Instructions CEREC Primemill
5 Installation and startup
5.4 Commissioning
67 19 681 D3439
D3439.201.01.01.02 09.2019
23
5.4.2
Supplied tools
5.4.2.1
Instruments
The following instruments are available for milling and grinding
purposes. When replacing instruments, ensure the permitted instrument
combinations are used (see "Permitted instrument combinations
[→ 41]").
5.4.2.2
Calibration pins
The calibration pins are used when calibrating the instrument sets (see
"Calibrating the unit [→ 38]").
5.4.2.3
Torque wrench
To insert or replace the instruments or calibration pins, use the
appropriate torque wrench. When doing so, pay attention to the
connection geometry of the torque wrench.
Instrument
REF
Usage
Color
Connection geometry of
the force transmission
Bur 2.5 ZrO2 CS
6713940
Milling of zirconium ox-
ide
(wet and dry)
Yellow
Square
Bur 1.0 CS
6713932
All-purpose milling
(wet and dry)
Black
Triangular
Bur 0.5 CS
6713924
All-purpose milling
(wet and dry)
Black
Diamond 1.4 CS
6714088
Grinding
White
Diamond 1.2 CS
6714070
Grinding
White
Diamond 1.0 CS
6714062
Grinding
White
Diamond 0.6 CS
6714054
Grinding
White
Calibration pin
Calibration
Light blue