DDS 7000 UNIT OVERVIEW
UNPACKING YOUR DDS 7000
Your DDS 7000 Dry Heat Sterilization System is delivered in two or more packages:
1. One large carton containing the DDS 7000 Dry Heat Sterilizer, Operating Instructions
and Accessory Kit.
2. One or more small cartons containing other accessories that may have been purchased.
WARNING!
This is a heavy piece of equipment. Please follow safe handling methods.
Examine each carton for any external damage. Be sure to check the corners of the box.
Immediately upon receipt open the large box, remove the protective packaging inserts and the sterilizer
from the carton. Carefully inspect the DDS 7000 for any damage.
Damage should be reported immediately to the transport carrier that delivered your unit as well as to the
Dentronix Sterilizer Repair Department 800-523-5944 (ext. 8925) or 330-916-8925 or e-mail tech@dentronix.
com or fax 330-645-8758.
CAUTION!
Read the Operating Instructions completely before operating the sterilizer
.
Set-up your DDS 7000 Sterilizer in a dry location according to the directions in the Operating
Instructions. Run your unit through a complete cycle with an empty chamber.
WARNING!
This is a heavy piece of equipment (96 pounds/43.64 Kg). Please make sure the counter
top will safely support the weight.
NOTE: RETAIN PACKAGING MATERIALS FOR FUTURE SHIPPING. The DDS 7000 should be
serviced by a professional on an annual basis, or when two consecutive spore strip test results
come back positive.
SERIAL NUMBER LABELS
Before proceeding further, check the specification label on the sides of the DDS 7000 to obtain
important information regarding model number, serial number and power requirements. Make
sure that this unit matches the power capabilities in your office regarding voltage, amperage
and cycles. Record the serial number for future reference. (See Introduction Section, Page 4)