3) Exhaust Louver replacement - Follow these instructions to ensure proper exhaust after
louver replacement.
a) Turn unit around, so that you are facing the back of the unit.
b) Remove exhaust stack and old louvers.
c) To insert new louvers:
• Start at the Bottom and work your way up.
• Insert one end first, with tabs facing away from you.
Then, gently flex the louver and insert the other end.
Important: Louvers on the left side of the chamber (intake)
must be positioned differently than louvers on the
right side (exhaust):
d) When all louvers are in place, turn unit around, so
that you are facing the door. (Looking into the
chamber, louvers on the left side open toward you.
Louvers on the right side open away from you.)
e) Reattach exhaust stack on the back of the unit.
f) Run a test cycle.
If you have any questions, contact the Dentronix Sterilizer
Repair Department.
Note: For replacement of intake louvers, repeat step C only.
The intake louvers are accessible from inside the chamber.
After installing new louvers, ensure that all louvers
move freely.
CLEANING: Outside of unit can be cleaned with warm soapy water. Wipe the inside of the sterilizer
chamber with a damp cloth. Any cleaner which leaves a residue may result in smoking or undesired
fumes during the sterilizing cycle.
Any other malfunction in operation should be reported to Dentronix Sterilizer Repair Department for further
instruction.