Once cool down is complete @ 104˚F (40˚C), the green “complete” LED lights up indicating that it is now safe to
remove the sterilized tools from the chamber (ABOVE).
WARNING!
DO NOT UNLOAD INSTRUMENTS UNLESS THE “CYCLE COMPLETE” LIGHT IS ON.
NOTE: The indicator lights will flash if there was a problem during any phase of the cycle. Refer to APPENDIX
II for error code explanations.
During cool down, the fan is drawing air through the HEPA Filter and into the chamber to cool the in-
struments. An audible tone will signal the completion of the cycle. The door must be opened to turn
this LED off and to begin a new cycle. If instruments are not immediately removed from the oven cham-
ber, they may reheat slightly from warm areas inside the chamber.
SPECIAL NOTE ON MOUTH MIRRORS
The DDS 7000 has been tested to safely sterilize good quality metal mouth mirrors. However,
depending on the age of your mirrors and their exposure to chemical disinfectants and cleaning
solutions, there may be moisture leakage behind the glass. This moisture will vaporize during the ster-
ilization cycle and cause shattering of the mirror. When sterilizing mirrors, wrap them in aluminum foil
to contain any glass fragments that may occur.
For best results, wipe mirror faces with a clean cloth after ultrasonic cleaning and before sterilization
to prevent fogging.