dentronix DDS 7000 Operating Instructions Manual Download Page 4

4

INTRODUCTION

Thank you for your purchase of the DDS 7000 Dry Heat Sterilization System. We appreciate your 
commitment to safe sterilization and effective patient care. The DDS 7000 Dry Heat Sterilizer is a 
convection type unit, designed for the sterilization of un-bagged orthodontic instruments that can 
withstand temperatures of up to 420ºF(215ºC). The DDS 7000 System is programmed to rap-
idly increase load temperature to 374ºF (190ºC), maintain this temperature for an effective and 
complete kill in three (3) minutes and then quickly cool the load for immediate handling. Optimum 
results are achieved with the Dry Heat method by uniform load distribution. Your DDS 7000 is 
equipped with a unique rack system for sterilizing hinged instruments. This design guarantees equal 
spacing of instruments and racks in the sterilization chamber and greatly simplifies loading, han-
dling, and dispensing of all instrumentation. Since no moisture, pressure, or chemicals are used in the 
DDS 7000 sterilization cycle, your valuable instruments are protected from rust, corrosion, and dulling 
of cutting edges.  The design includes a safety interlock switch on the door to prevent access to the 
chamber until the load has been properly cooled. Dentronix is confident that you will find this sys-
tem easy to integrate into your office procedures, as well as simple to use by all staff members. The 
following operating instructions have been written to familiarize you with the principles of Dry Heat 
Sterilization and the specific operating procedures for the DDS 7000. It is imperative that you read 
these operating procedures carefully before operating your unit. 

NOTE: The DDS 7000 Sterilizer has been manufactured for specific and safe operating procedures 
as contained in this manual.  

CAUTION!

Never use the DDS 7000 for other uses. i.e. sterilization of bagged instruments or 
use of cassettes, as a warming chamber or heating device, for the preparation of 
food or other items. If you are unsure of your unit’s correct usage, please contact 
your sales representative or the Dentronix Customer Service Hotline.

Questions?  Please call the Dentronix Sterilizer Repair Hotline – 800-523-5944 (ext. 
8925)  in U.S. and Canada or 330-916-8925 or e-mail [email protected] or fax 330-
645-8758 so that we can help you maximize the full capabilities of the DDS 7000 System.

Please record your Serial Number here for more efficient customer service:

Summary of Contents for DDS 7000

Page 1: ...DDS 7000 Dry Heat Sterilization System 115V 230V Operating Instructions...

Page 2: ...2...

Page 3: ...INTERLOCK 12 FIRST TIME USE CYCLE TEST 12 STERILIZING YOUR INSTRUMENTS 12 INSTRUMENT HANDLING 13 PLIER RACKS 14 LOADING YOUR DDS 7000 PLIER RACKS 14 LOADING THE STERILIZATION CHAMBER 16 OPERATION 17 S...

Page 4: ...ncludes a safety interlock switch on the door to prevent access to the chamber until the load has been properly cooled Dentronix is confident that you will find this sys tem easy to integrate into you...

Page 5: ...s on pages 14 15 Same Sterilization Temperature 374 F 9 F 190 C 5 C Same Controls Displays Standby On switch Same Cycle Start switch Same Lighted indicators for Warm up Sterilize Cool Down and Cycle C...

Page 6: ...UNIT OVERVIEW Figure 1 Outside view of the DDS 7000 Louvers Door Switch COM Port Rack Channels Standby On Door Lock Digital Display Cycle Start Switch Cycle Indicator Lights Magnetic HEPA Filter Cove...

Page 7: ...925 or e mail tech dentronix com or fax 330 645 8758 CAUTION Read the Operating Instructions completely before operating the sterilizer Set up your DDS 7000 Sterilizer in a dry location according to t...

Page 8: ...teri als Leave at least 4 10 2 cm side clearance between cabinets and walls to allow unrestricted passage of cool down intake air Failure to allow a four inch side clearance could result in a consider...

Page 9: ...on of the DDS 7000 in a small room less than 12 x 12 3 66m x 3 66m b Installation of the DDS 7000 under cabinets or built into cabinets Two styles of vent adapters are available See Figure 4 Below Typ...

Page 10: ...operate at 50 60 Hz STERILIZER CONTROLS The DDS 7000 is an automatic sterilizing system Sterilization time and temperature have been pre set at the factory providing for optimum performance 1 Standby...

Page 11: ...ature 4 Cycle Complete this Green LED will illuminate at the completion of the cool down phase and indicates that instruments are sterilized and ready to unload It is also your indication that the DDS...

Page 12: ...n Error Code will display notifying the operator of an interruption Toggling the Standby On switch from On to Standby then back to On will clear the error message To restart the cycle press Cycle Star...

Page 13: ...TION ITEMS THAT ARE LABELED AUTOCLAVABLE ARE NOT NECESSARILY DRY HEAT STERILIZ ABLE IF IN DOUBT ABOUT SPECIFIC PRODUCTS CONTACT THE MANUFACTURER CAUTION REMOVE ALL VINYL GRIPS FROM PLIERS BEFORE STERI...

Page 14: ...instruments are dropped sprayed or mishandled after sterilization the instruments must be pre cleaned as recommended above Never return the items to the sterilizer without first cleaning them WARNING...

Page 15: ...igned to provide a predictable location and spacing for the instrument load so that optimum exposure can be achieved Pliers should only be loaded in the notches of the rack see Figure 6 Never place in...

Page 16: ...two 2 combo racks each holding five 5 Mathieu Hemostats or other hand instruments Full Tray Only One 1 Horizontal Tray capable of holding up to six 6 orthodontic pliers or up to 12 single handle hand...

Page 17: ...two Plier Racks Figure 7C Improperly loaded Half Tray with Figure7D Close up of improperly loaded Plier Rack The Plier Rack is not in the proper half tray Instruments should be placed so rack channel...

Page 18: ...bient temperature No LEDs are lit at this time Press the Cycle Start switch to initiate a cycle The cycle is preprogrammed and cannot be altered by the user Figure 8C Figure 8D Once the Cycle Start sw...

Page 19: ...s from the lower right of the display Once the operating temperature of 374 F 190 C is reached the DDS 7000 begins the sterilize exposure phase The amber Sterilize LED lights up During this phase the...

Page 20: ...this LED off and to begin a new cycle If instruments are not immediately removed from the oven cham ber they may reheat slightly from warm areas inside the chamber SPECIAL NOTE ON MOUTH MIRRORS The DD...

Page 21: ...on where in the sterilization cycle the interruption occurred the contents of the chamber could be extremely hot Appropriate protective equipment should be worn when handling the contents of the chamb...

Page 22: ...the DSSH 1 Spore Strip Holder supplied Use of the Spore Strip Holder is recommended in order to keep the biological indicator from moving around the chamber which may result in false test results 1 Pl...

Page 23: ...ith correct amp fuse See General and Technical Specifications Page 5 2 HEPA Filter located below unit Remove magnetic strip and replace filter when dirty or if an error code is indicated during cool d...

Page 24: ...king into the chamber louvers on the left side open toward you Louvers on the right side open away from you e Reattach exhaust stack on the back of the unit f Run a test cycle If you have any question...

Page 25: ...the Standby On switch to the Standby position 2 Open the door releasing the Door Switch 3 Press and hold the Cycle Start Switch until the display shows 4 red dots 5 seconds The display will go blank...

Page 26: ...ure is to be used if you suspect that the sterilizer is not working properly The Self Test will cause each of the systems to operate in a short period of time to assure that all systems are working If...

Page 27: ...rt printout for hand written user information See the following page for a representative printout of a PASS and FAIL cycle Since the device is not set up to automatically print the date you will have...

Page 28: ...Temp 104F Time 18 15 Status Pass COMPLETE Time 44 21 Status Pass COMMENTS Dentronix DDS 7000 Dry Heat Sterilizer a subsidiary of Coltene Whaledent Copyright 2004 All Rights Reserved FIRMWARE VERSION...

Page 29: ...tons and packaging materials are available from Dentronix at a nominal fee Loaners are available at no charge if unit to be repaired is under warranty After one year from date of sale loaners are avai...

Page 30: ...QUALIFIE CAUTION HOT SURFACE PRUDENCE SURFACE CHAUDE AVERTIR LA SURFACE CHAUDE CAUTIONS and WARNINGS IMPORTANT NOTICE AVERTIR LA NOTIFICATION IMPORTANTE PROTECTIVE GROUND SOL PROTECTIF Appendix I Sym...

Page 31: ...ameter fails before pre cooldown phase STANDBY ON switch is switched from ON to STANDBY Cycle fails to reach warm up conditions E 07 can occur if Temperature Overload Sensor is tripped See page 20 for...

Page 32: ...hoga Falls OH 44223 USA Tel USA Canada 1 800 523 5944 1 330 916 7300 Fax 1 330 916 7333 info dentronix com P N 30003185B 06 22 15 Coltene Whaledent GmbH Co KG Raiffeisenstrasse 30 89129 Langenau Germa...