Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
13
Remarks
When“
”changed to “E1”means heating malfunction
,
please check your
heating needle.
LCD display
Shows preset temperature or
real-time temperature. Middle shows
the heating needle status:
“
√
”heating needle connected,
“×”heating needle broken or no
needle.
Heating Tips
Type F,FM,M,ML 4 pcs in total
Notice
:
Rotary, twist and strong
tug is forbidden
Charging base
Place and charging the heating pen
1)Use adapter as 7.5V 2.5A
2)After the adapter power is
connected, the charging lamp
shows the red and green alternately
to indicate that the battery is to be
charged (not charged)
3)Place the heating pen onto the
charging base with the button
inward
(
match well the charging
electrode
)
,the charging lamp is red and bright 4)The power is full when the
charging lamp is green.
Remarks
Charging lamp
Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
14
The charging base is not waterproof, the liquid contact electrode is
forbidden, and the conductor is forbidden to contact the charging
thimble.
SECTION 6: INSTRUCTION OF OPERATION
It’s important to keep following points to use this system efficiently and to
avoid any mechanical trouble or dangerous condition.
Before using GP Obturation System, the dentists need to make good root canal
preparation with endo motor.
Requirement for good root canal preparation:
1. All the root canal should be made as straight as possible.
2. Root canals especially small collateral root canals should be cleaned as clean
as possible.
3. The dentists must possess root canal cleaner like sodium hypochlorite and
had better get a ultrasonic root canal cleaning device to make root canal clean
and smooth enough for obturation.
During Operation of vertical condensation with obturation pen:
1. The heat tips should be 4-5mm far from apex point of root canal.
2. The heat tips should be fixed with white slide caliper provided after
determining the length of root canal.
3. The gutta percha points (cone) should be 0.5-1mm shorter than the length
of root canal.
4 Root canal perforation should be avoided when heating. This is very very
important.
5. All heat tips are easily broken although the quality of heat tips is very good!
Heat tips are consumable!
When the dentists use obturation pen, they should be very tender and soft.
During Operation of backfilling with obturation gun:
Injection needles are also very fragile because it is made of silver!