Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
25
The [DY-GP Plus] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [DY-GP Plus] should assure
that it is used in such an environment
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The [DY-GP plus] uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class [B]
The [DY-GP Plus] is suitable for
use in all establishments other
than domestic, and may be used in
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes,
provided the following warning is
heeded:
Warning
: This equipment/system
is intended for use by healthcare
professionals only. This
equipment/ system may cause
radio interference or may disrupt
the operation of nearby equipment.
It may be necessary to take
mitigation measures, such as
re-orienting or relocating the
[DY-GP Plus] or shielding the
location.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies
Table 2
Guidance and manufacturer’s declaration - electromagnetic emissions
The [DY-GP Plus] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [DY-GP Plus] should assure
Rev. 02/01/18 VER SMS-DYGP-PLUS-FREEFILL20170828-EN
26
that it is used in such an environment
Immunity Test
IEC 60601
Test level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
The electrical fast
transient burst (EFT) is
generated by the
switching of inductive
loads. Separation
between the equipment
and other loads shall be
considered before
installation. Mains filter is
required, if necessary.
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
±1 kV line(s)
to line(s)
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in
U
T
)
for 0.5 cycle
40% U
T
(60% dip in
U
T
)
for 5 cycle
70% U
T
<5% U
T
(>95% dip in
U
T
)
for 0.5 cycle
40% U
T
(60% dip in
U
T
)
for 5 cycle
70% U
T
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the [DY-GP Plus]
requires continued
operation during power
mains interruptions, it is
recommended the
[DY-GP Plus] be powered
from an uninterruptible