7. After use, extended press POWER button for switch off, if necessary. 
8. Remove the heat tips, clean the device and related accessories for next use. 
 

 

Remarks

 

During the process of heating, extended press the button of 10S can 

switch off heating automatically. 

Warning Tone: 1 sound for normal operation, 2 for reminder, 3 for 

incorrect operation or invalid operation.   

  For the sake of safety, do not allow heat tips to stay in the root canal 

for more than 5S during the working condition. 

 

C. Charging System   

 
1. Checking the charging state, when the battery is charging, the indicator light 
shows red color, when the battery is fully charged, the indicator light shows 
green. 
2. When the battery is charging or stop using, please do not put it near metal 
in order to avoid the short circuit of the machine. 
3. Please do not use other adapter if the adapter is damaged, please contact 
with distributor or manufacturer for solution!   
 

 

Remarks

 

 

※ 

1) Heat tips, injection needles and heat insulator

 

must be sterilized 

before use;

 

※ 

2) Because of the rapid warming and cooling characteristics of the 

FreeFill pen, special care is needed to avoid any accidents by burning; 

※ 

3) The FreeFill gun can enter the root canal after heating, and pay 

attention to prevent it from being burned by injection needle; 

※ 

4) The FreeFill gun is forbidden to pull the trigger when gutta percha 

bar is inserted, but the heating function is not turned on, so as to avoid 
mechanical trouble. 

 

     

Rev. 02/01/18      VER SMS-DYGP-PLUS-FREEFILL20170828-EN

 

 

20 

 

SECTION 7: APPLICATION OF ROOT CANNAL FILLING 

First-Phase Preparation 

1) Gutta percha point (Master cone) 
Select the appropriate gutta percha point as a master cone which has same 
taper and size with prepared root canal. Shorten 0.5-1 mm

 

of the point than 

working length to avoid overfilling and to secure enough space to fill 
additional point. 
2. Hand plugger 
Prepare 2-3 hand plungers to compact the filled gutta percha.   
3. Pen tips 
Select the appropriate pen tip and sign the working length by using callipers, 
The signed working length should be 5-7 mm short than the length of root 
canal and make sure the pen tip size will not touch the root canal. 
4. Pen tips 
Please select the appropriate size of pen tip according to the thickness at the 
2/3 position of root canal coronal. 
 

A.

 

Vertical Condensation/Down Pack 
 

1. Dry the root canal, apply a thin layer of root canal sealer on the root canal 
wall, and put the GP cone in position. 
2. Confirm the working length, warm the selected pen tip and cut off excess GP 
cone.  
3. Put the gutta percha cone inside the root canal and then push it slowly into 

the apex by using warmed pen tip until the pen tip is 5-7 mm from the apex 
then stop warming the pen tip. (The whole process should be finished in 7-9 
seconds) 

 
4. Softly press the gutta percha cone for 10 seconds, warm the pen for 1 
second again, and take out residual gutta percha cone. Compact the warmed 
gutta percha by using the hand squeezing device. 
5. If the root canals need to place the post pin, the backfilling process can be 
omitted, leaving enough space for the post pin. If not, backing filling it with 
obturation gun.   

Summary of Contents for DY-GP Plus

Page 1: ...FREEFILL20170828 EN USER S MANUAL DY GP Plus Cordless GP Obturation System The unit must be installed by a qualified engineer Only for user by dental professionals Read this operation manual carefully before installation or operation ...

Page 2: ...need to notice the user in advance CONTACT INFORMATION The device is manufactured by DENJOY DENTAL CO LTD Address F4 Building A4 Lugu Medical Device Park No 229 Guyuan Road Changsha 410205 P R China Website www denjoy cn E mail denjoy denjoy cn Phone 86 731 85126543 Fax 86 731 88885317 Authorized European Representative Company name LANDLINK GMBH Address DORFSTRASSE 2 4 79312 EMMENDINGEN GERMANY P...

Page 3: ... for injection needle z Faster power charging shorter charging time z Smart power saving mode offers longer operation z Light and more comfortable more ergonomic design z All parts which touched the patients can be autoclaved z Two in one system makes filling easier and more convenient z Digital display of temperature regulation it is more convenient z Special mode for changing injection needle an...

Page 4: ... be used to cut off excessive gutta percha point SECTION 3 SYMBOL The following symbols may appear in this manual on the label or on it s accessories Some of the symbols represent standards and compliances associated with the device and its use Consult accompanying documents Caution Authorized Representative in the European Community Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 6 CE Mark con...

Page 5: ... 18 VER SMS DYGP PLUS FREEFILL20170828 EN 8 Injection Device FreeFill Obturation Gun FreeFill Gun Injection device For backfilling the melted gutta percha Plugger Pushing gutta percha outside of the needle Status Description of the Plugger Gutta percha is pushed inside of the injection needle please start heating and backfilling when set temperature reached Gutta percha is used up please start the...

Page 6: ...battery Full power Cleaning status constant temperature is 100 quit cleaning function by press any button Energy saving status constant temperature is 120 quilt this function when hold the inductive switch or press button S this mode can be shut off manually Remarks When changed to E1 means heating malfunction E2 means warming up malfunction With the device off extended press button POWER until 2 ...

Page 7: ...s the red and green alternately to indicate that the battery is to be charged not charged Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 12 3 Place the injection device onto the charging base with the trigger inward match well the charging electrode the charging lamp is red and bright 4 The power is full when the charging lamp is green Remarks The charging base is not waterproof the liquid con...

Page 8: ...g points to use this system efficiently and to avoid any mechanical trouble or dangerous condition Before using GP Obturation System the dentists need to make good root canal preparation with endo motor Requirement for good root canal preparation 1 All the root canal should be made as straight as possible 2 Root canals especially small collateral root canals should be cleaned as clean as possible ...

Page 9: ...t s provides easy and fast filling but it s recommended to use the pen to fill the apical part at first to avoid any overfilling or under filling and the gun can be used for filling the remained part Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 16 of the canal 1 Select the appropriate injection needle and attach it to the gun Tighten firmly using the accessory tools provided Then put the bat...

Page 10: ...on needle to avoid burning Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 18 B FreeFill Obturation Pen The pen is designed for vertical condensation or down packing with warmed gutta percha in the canal Besides it can be used to cut off excessive gutta percha point 1 Dry root canal Apply a thin layer of root canal sealer to the root canal wall and gutta percha points then insert the gutta perc...

Page 11: ... trouble Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 20 SECTION 7 APPLICATION OF ROOT CANNAL FILLING First Phase Preparation 1 Gutta percha point Master cone Select the appropriate gutta percha point as a master cone which has same taper and size with prepared root canal Shorten 0 5 1 mm of the point than working length to avoid overfilling and to secure enough space to fill additional poin...

Page 12: ...y case If you want to clean the pen tip during operation Heating state please use dried towel do not use wet cleaning solution Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 22 SECTION 9 TROUBLE SHOOTING TROUBLES CHECK Can not be recharged Check the power indicator Check that the adapter and charging base are connected properly Check that the battery and charging base are connected properly Ch...

Page 13: ... within warranty period Do not modify and disassemble the device This instrument described below has been fully inspected and confronts to the current products specification The control part is guaranteed for its designated use against original defects in materials and workmanship for a period of 12 months from date of purchase The guarantee for spare parts is 6 months from date of purchase Produc...

Page 14: ... environment specified below The customer or the user of the DY GP Plus should assure Rev 02 01 18 VER SMS DYGP PLUS FREEFILL20170828 EN 26 that it is used in such an environment Immunity Test IEC 60601 Test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile...

Page 15: ...diated RF IEC 61000 4 3 80MHz to 2 5GHz d 1 2 d 1 2 80MHz to 800MHz d 2 3 800MHz to 2 5GHz where p is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in ...

Page 16: ...mitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 80MHz to 800MHz d 1 2 800MHz to 2 5GHz d 2 3 0 01 0 12 0 23 0 1 0 38 0 73 1 1 2 2 3 10 3 8 7 3 100 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where...

Reviews: