14
Figure 4-1. CUP 15 Grounding Receptacles
Power Requirements
The CUP 15 requires a dedicated 230V, 50Hz / 60 Hz, 30 amp circuit. Please ensure
that all receptacles (NEMA L6-30R) match the required power ratings (see
Figure
4-1
).
NOTE:
Always plug the system into correctly grounded receptacles.
Figure 4-2
shows properly wired grounding straps and their placement.
NOTE:
Grounding straps should be wired by a licensed electrician.
Figure 4-2. System Grounding Straps