8
3.
USER
INTERFACES
The user interface is composed of an
LCD
graphic back - enlighten display
and start and set keys
of the instrument. With a battery power supplying without pressing any key, the back- lighting
switches off after 1 minute. To activate it again, press one key. With external power supply, the
back-lighting is always activated.
To switch the instrument on and off, press
ON/OFF
key. When you switch it on, the logo and the
model of the instrument will be visualize, for some seconds, then you can go to the main visualiza-
tion.
3.1
THE DISPLAY
WBGT Index
2008/11/28 08:00:00
Log 00 00:00:00
Tn 15.6 °C
Tg 20.2 °C
T 20.2 °C
WBGT(in) 17.0 °C
WBGT(out) 17.0 °C
The first line visualizes the
charge condition of the battery
, the
second line indicates the current
date and the current hour
. If logging function is activated, the third line indicates the current log-
ging number and the spent time from the logging start.
The
detected sizes
are:
Tn:
wet bulb with natural ventilation temperature
Tg:
globe thermometer temperature
T:
ambient temperature
WBGT (in):
WBGT index in absence of direct solar irradiation
WBGT (out):
WBGT index in presence of direct solar irradiation
Summary of Contents for HD32.2
Page 2: ...2 WBGT Wet Bulb Globe Temperature Meter HD32 2...
Page 39: ...39 12 PRINTING OF THE MEASUREMENT REPORT Norm ISO 7243 Norm ISO 7243...
Page 40: ...40 Warm Environments Norm ISO 7243...
Page 41: ...41 Warm Environments Norm ISO 7243 Norm ISO 7243...
Page 42: ...42 Warm Environments Norm ISO 7243...
Page 43: ...43 Warm Environments Norm ISO 7243...
Page 44: ...44 Warm Environments Norm ISO 7243...
Page 49: ...49 NOTES...
Page 50: ...50 NOTES...