10.2  ON LINE INSTRUMENT PROBES AND MODULES TECHNICAL DATA 

 

10.2.1 Wind speed measurement probes 

 

Hot-wire probes: AP471 S1 - AP471 S2 - AP471 S3 - AP471 S4 - AP471 S5

 

 

 

AP471 S1 - AP471 S3

 

AP471 S2

 

AP471 S4 
AP471 S5

 

Type of measurements

 

Wind speed, calculated flow rate, air temperature 

Type of sensor

 

 

 

  Speed 

NTC thermistor 

Omnidirectional NTC thermistor

  Temperature 

NTC thermistor 

NTC thermistor 

Measurement range

 

 

 

 Speed 

0.05…40m/s 

0.05…5m/s 

 Temperature 

-30…+110°C 

-30…+110°C  0…80°C 

Measurement resolution:

 

 

 

 Speed 

0.01 m/s 

0.1 km/h 

1 ft/min 

0.1 mph 
0.1 knot 

 Temperature 

0.1°C  

Measurement accuracy

 

 

 

  Speed 

±0.1 m/s   (0.05…0.99 m/s) 

±0.02m/s    (0.05…0.99 m/s) 

 

±0.2 m/s  (1.00…9.99 m/s) 

±0.1m/s      (1.00…5.00 m/s) 

 

±0.6 m/s    (10.00…40.0 m/s) 

 

  Temperature 

±0.4°C      (-30…+110°C) 

±0.4°C        (-30…+110°C) 

Minimum speed 

0.05 m/s 

Air temperature 
compensation 

0…80°C 

Battery life 

Approx. 20 hours @ 20 m/s with 
alkaline batteries 

Approx. 30 hours @ 5 m/s with 
alkaline batteries 

Unit of Measurement

 

 

  Speed 

m/s – km/h – ft/min – mph – knot 

  Flow rate 

l/s - m

3

/s - m

3

/min - m

3

/h - ft

3

/s -  ft

3

/min 

Pipeline section for flow rate 
calculation 

0.0001…1.9999 m

2

 

Cable length 

~2m 

 
 

Summary of Contents for HD2303.0

Page 1: ...ation written in this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manua...

Page 2: ...2 Anemometer HD2303 0...

Page 3: ...the current value 8 UNIT MENU key it allows selection of the unit of measurement when pressed together with the DATA key it allows to open the menu 9 STD BY during normal operation it sets the hot wi...

Page 4: ...E PROBES 21 4 4 1 Temperature measurement 21 4 4 2 Connecting the TP47 connector for the 4 wire Pt100 and 2 wire Pt1000 probes 21 4 4 3 Direct connection of 4 wire Pt100 sensors 22 5 WARNINGS 23 6 INS...

Page 5: ...or Pt1000 The probes are fitted with the automatic detection module with the factory calibration settings already being memorized inside The units of measurement for the measurable quantities are for...

Page 6: ...FAST or SLOW is displayed then the instrument carries on with the standard measurement MIN cm min s 2 3 m m k AUTO OFF the AutoPowerOff function can be disabled by simultaneously pressing this key and...

Page 7: ...TA calculation function When the instrument is off the previously memorized data are cleared Upon turning on the instrument automatically starts memorizing the MAX MIN and AVG values To reset the prev...

Page 8: ...inch2 this parameter defines a duct s section area for flow rate calculation 2 The filter FAST or SLOW used during measurements 3 Probe type the menu is opened by pressing simultaneously DATA UNIT the...

Page 9: ...go on to the next item push DATA ENTER push CLR ESC to delete the modification push UNIT MENU key again to quit 3 Probe type the PRBE_TYPE message is displayed in the comment line The main line in the...

Page 10: ...te Some also measure air temperature The measurement principles used are the following hot wire for the AP471 series vane for the AP472 series On request the probes of the AP471 and AP472 series can b...

Page 11: ...the SECT m2 menu item press STD BY to change the unit of measurement from m2 to inch2 using the 5 and arrows enter the desired value press DATA ENTER to confirm the modification and go onto the next...

Page 12: ...m s in any direction of the air flow incident on the probe The AP471 S4 probe is also fitted with support base and sensor protection The AP471 S5 is identical to the AP471 S4 but instead of a base it...

Page 13: ...ck it in measurement condition completely low or rest condition completely high To reduce the space occupied when not used the AP471 S4 and AP471 S5 are supplied with a protection cylinder that can be...

Page 14: ...3 Introduce the probe in the air flow being measured maintaining the arrow at the top of the probe parallel to the flow as indicated in the figures NOTE the probe should be maintained orthogonal to t...

Page 15: ...mnidirectional probes that have an uncovered sensor could render the probe unusable After measurement the sensor set on the probe head must be protected with the supplied metallic screen or threaded c...

Page 16: ...16 MIN 350 MAX 1020 120 AP471 S4 AP471 S5...

Page 17: ...Thermocouple K 16 The temperature limit refers to the probe head where the vane and temperature sensors are located and not to the handle cable and telescopic rod that can be subjected to maximum temp...

Page 18: ...ided in the introductory paragraphs of this chapter 4 3 3 Care and maintenance of the probes The probe performance mainly at low speeds largely depends on the very slow friction of the vane on its own...

Page 19: ...19 4 3 4 Dimensions 1 100 3 60 1 3 16 1 2 3 AP472 S1 AP472 S2 AP472 S4...

Page 20: ...on shaft diam 16mm In order to screw the handle off 3 hold the probe tight on point 1 Screw the shaft top AP471S1 23 6 on the screw 2 Other extension shafts AP471S1 23 6 can be added The handle 3 or t...

Page 21: ...he temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip probe In order to perform a correct contact measurement the measurement surface mu...

Page 22: ...ct 7 Once the welding operation is complete close the two shells 8 insert the gasket in the module 9 screw the fairlead and the ring Make sure the cable is not twisted while you are screwing the fairl...

Page 23: ...0 C avoid violent blows or thermal shocks to Platinum temperature probes as this could cause irreparable damage 9 To obtain reliable measurements temperature variations that are too rapid must be avoi...

Page 24: ...t the probe is measuring a value exceeding the measuring range PLS_EXIT FUNC RES_FOR_FACT ONLY please exit using ESC function reserved to factory calibration PRBE_SER serial number of the connected pr...

Page 25: ...nstrument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 scre...

Page 26: ...safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product...

Page 27: ...ent off 20 A Connections Input for probes 8 pole male DIN45326 connector Unit of Measurement m s km h ft min mph knot l s m3 s m3 min m3 h ft3 s ft3 min C F Measurement of temperature by Instrument Pt...

Page 28: ...ure 30 110 C 30 110 C 0 80 C Measurement resolution Speed 0 01 m s 0 1 km h 1 ft min 0 1 mph 0 1 knot Temperature 0 1 C Measurement accuracy Speed 0 1 m s 0 05 0 99 m s 0 02m s 0 05 0 99 m s 0 2 m s 1...

Page 29: ...8 20 10 50 Temperature C 25 80 25 80 25 80 30 120 25 80 30 120 Resolution Speed 0 01 m s 0 1 km h 1 ft min 0 1 mph 0 1 knot Temperature 0 1 C 0 1 C 0 1 C Accuracy Speed 0 2 m s 1 5 f s 0 2m s 1 5 f s...

Page 30: ...C 400 C TP475A 0 Air 50 C 250 C 0 3 C 50 C 250 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP472I 10 Immersion 50 C 400 C 0 3 C 50 C 350 C 0 4 C 350 C 400 C TP875 Globe thermom...

Page 31: ...c probe measuring range 0 05 5m s Cable length 2 metres VANE PROBES AP472 S1 Vane probe with thermocouple 100mm Speed from 0 6 to 25m s temperature from 25 to 80 C Cable length 2 metres AP472 S2 Vane...

Page 32: ...2 metres TP472I 5 Pt100 sensor immersion probe Stem 6 mm length 500 mm Cable length 2 metres TP472I 10 Pt100 sensor immersion probe Stem 6 mm length 1000 mm Cable length 2 metres TP875 Globe thermomet...

Page 33: ...e misure ai campioni internazionali e nazionali delle unit del SIT garantita da una catena di riferibilit ininterrotta che ha origine dalla taratura dei campioni di laboratorio presso l Istituto Prima...

Page 34: ...electronic devices with the following symbol cannot be disposed in the public dumps According to the Directive UE 2002 96 EC the European users of electric and electronic devices are allowed to give...

Reviews: