![Delta OHM HD2303.0 Manual Download Page 27](http://html1.mh-extra.com/html/delta-ohm/hd2303-0/hd2303-0_manual_2478302027.webp)
- -
27
10.
TECHNICAL
CHARACTERISTICS
10.1
T
ECHNICAL INFORMATION ON THE
A
NEMOMETER
Instrument
Dimensions (Length x Width x Height)
140 x 88 x 38 mm
Weight
160 g (complete with batteries)
Material ABS
Display
2x4½ digits plus symbols
Visible area: 52 x 42 mm
Operating conditions
Operating temperature
-5
÷
50°C
Warehouse temperature
-25
÷
65°C
Working relative humidity
0
÷
90%RH without condensation
Protection degree of the case
IP67
Power
Batteries
3 1.5V type AA batteries
Autonomy
(*)
200 hours with 1800mAh alkaline
batteries
Power absorbed with instrument off
< 20
μ
A
Connections
Input for probes
8-pole male DIN45326 connector
Unit of Measurement
m/s-km/h-ft/min-mph-knot
l/s-m
3
/s-m
3
/min-m
3
/h-ft
3
/s-ft
3
/min
°C - °F;
Measurement of temperature by Instrument
Pt100 measurement range
-200
÷
+650 °C
Pt1000 measurement range
-200
÷
+650°C
Resolution
0.1 °C
Accuracy
±
0.1 °C
Drift after 1 year
0.1 °C/year
EMC standard regulations
Security
EN61000-4-2, EN61010-1 level 3
Electrostatic discharge
EN61000-4-2 level 3
Electric fast transients
EN61000-4-4 level 3,
EN61000-4-5 level 3
Voltage variations
EN61000-4-11
Electromagnetic interference susceptibility IEC1000-4-3
Electromagnetic interference emission
EN55020 class B
(*)
Valid for all probes except hot-wire probes. For their autonomy see the characteristics below.