HD2107
- 17 -
V2.3
Sensor
Board connection
Jumper to close
Ni1000
JP4
4
JP3
3
JP1
1
JP2
2
Ni1000
2 wires
JP3
Ensure the welds are clean and perfect. Once the welding operation is complete, close the two
shells, insert the gasket in the module, and screw the fairlead and the ring. At the other end of the
module, enter the ring with the O-Ring as indicated in the picture.
Make sure the cable is not twisted while you are screwing the fairlead. Now the probe is ready.
Direct connection of 4 wire Pt100 sensors
4 wire Pt100
sensors
can
be soldered directly
to the pins of the free female connector with-
out making use of the TP47 board. The 4
wires of the Pt100 sensors have to be soldered
as indicated in the figure on the left.
In order to use this type of probe it is neces-
sary to set up the menu item “Probe Type” as
described at page 10.
The P100 probe is recognized upon turning on
the instrument: connect the probe when the in-
strument is switched off and then turn it on.
Sensor
Direct soldering to the connector
Pt100 4
wires
View of the soldering side
of the free female connector