9
Non-selectable items:
o
SCALE NAME (top line)
o
DRY/WET bar
o
The MEMORY status (top right corner)
Selectable items:
o
MC reading (large text)
o
JOB (current job)
o
ALARM (current alarm setting)
o
MODE (current measurement mode)
o
EXIT (leave MC READ Mode)
NOTE:
The default selection on this screen is MC reading, which places the meter in measurement
mode. Pressing the SELECT key causes a single reading to be taken (Triggered mode) or initiates a
reading stream (Continuous mode).
READ MODE – CONTINUOUS/TRIGGERED
The meter has two measurement modes: Triggered and Continuous. Choosing the MODE option and
pressing the SELECT key will change the measurement mode.
o
Triggered mode causes a single reading to be taken every time the SELECT button is pressed.
Each reading is stored in memory. The last measurement remains displayed on the screen.
o
Continuous mode starts by pressing the SELECT key. The meter keeps taking MC readings in
this mode. These readings are NOT stored until the SELECT key is pressed again. To exit
continuous mode WITHOUT STORING the current reading, press any directional key (UP,
DOWN, RIGHT, LEFT). To exit continuous mode and STORE the current reading, press the
SELECT key. Exiting continuous mode leaves the last reading on the screen. To resume
continuous mode, press the SELECT key again. Use the directional keys to choose a different
option.