represent the relative dryness or wetness of the material on a scale of 5 to 60. In order to 
accurately interpret your readings, first establish a baseline reading by taking a reading in an 
area you know to be unaffected, then take a reading in the affected area and compare the 
two readings. 

 

MEASURING RELATIVE HUMIDITY IN CONCRETE FLOOR 
SLABS IN ACCORDANCE WITH ASTM STANDARD F2170: 

 

Select test locations to provide information about moisture distribution across the entire concrete 
slab, especially areas of potentially high moisture, or if the slab is thick (>4 in.) and air circulation 
is poor. For slabs on-grade and below-grade, include a test location within 3ft of each exterior 
wall. Perform minimum three tests for the first 1000 ft sq and at least one additional test for each 
additional 1000 ft sq.  
 

 

1.  Drill a 5/8” hole to a depth of 40% of the slab thickness using a rotary hammer drill. Use a 

drill bit specially designed for rotary drills and with 3 or 4 cutting edges to insure a 
smooth, round hole. Surface preparation is not required. 

 

2.  Vacuum the hole thoroughly, then use a wire brush to clean and loosen any concrete 

remaining in the hole. Vacuum again and repeat the process a second time. 

 

3.  Insert the yellow sleeve  (sold separately)  into the hole. use silicone (or other water-

resistant sealant) to seal the interface between the collar of the sleeve and the concrete 
surface

Use a hammer to insure the sleeve is fully inserted into the hole and sealed to 

the concrete. 

 

4.  Insert the protective yellow cap into the sleeve and let the hole acclimate for 72h.  

 

5.  Remove the cap and immediately insert the sensor to full hole depth. Wait a minimum of 

1h and then take RH readings by connecting one end of the RH/T-C1 cable to the top of 
the meter and the other end to the exposed connector on the sensor housing. Check to 
ensure the meter’s displayed RH reading is no long changing.  The reading must not 
change more than 1% RH over 1 hour.  If the reading does change, allow more time for 
acclimation. 

 

6.  Alternately, insert the RH sensor into the sleeve immediately after setting the sleeve. 

Readings can then be taken as soon as the hole has acclimated.  
 

The complete ASTM document is available  at: 

http://www.astm.org/Standards/F2170.htm

 

 

 
TESTING CONCRETE SLABS FOR FLOORING APPLICATIONS

 

 

Moisture meters are an effective tool to check moisture in concrete. They can tell you where there 
may be excess moisture and help determine if you need to conduct further testing. 
 
It is important to test both the surface and mid-section of the slab, especially if the slab is on or 
below grade.  This will help determine if there is continuous moisture migration toward the 
surface. If this condition exists, the moisture movement may be so slow that once it reaches the 
surface, moisture evaporates and causes a “dry” reading when a surface test is made.  
 
However, if a sub-surface test is made, the meter may read “wet” indicating the presence of 
moisture. When the slab is covered and the upward movement of moisture continues, moisture 
will move into a hygroscopic (wood) floor, or build-up pressure under a non-breathing synthetic 
floor, causing delamination.   
 

Summary of Contents for QuickNav

Page 1: ...QuickNav MOISTURE METER Owner s Manual Version 1 0 Delmhorst Instrument Co Rev 1 June 2011...

Page 2: ...READINGS 5 6 TO TAKE DRYWALL READINGS 6 TO TAKE SCAN READINGS 7 TO TAKE TEMP RH READINGS 7 TO SET THE ALARM IN WOOD DRYWALL SCAN MODE 8 MAIN MENU OPTIONS 8 TAKING A READING PRACTICAL APPLICATIONS 9 TE...

Page 3: ...the range of 0 100 with accuracy of 2 over 10 90 Measures temperature over the range of 40 F 255 F with accuracy of 1 8 F over 4 F to 158F range of 40 C 124 C with accuracy of 1 C over 20 to 70 C Calc...

Page 4: ...Open the battery compartment by sliding the lid back while pressing on the release indent 2 Ensure correct polarity and push the battery in flush with the bottom board until the connectors snap togeth...

Page 5: ...ice The next screen is the MAIN MENU Use the keys to select the desired function mode and then press SELECT to activate the function Keypad Layout METER USE TO TAKEWOOD READINGS 1 From the Main Menu u...

Page 6: ...the key to highlight DRYWALL Then use the SELECT key to enter the pin mode 2 The meter will display the DRYWALL screen as shown below 3 Push the contact pins into the drywall Any Delmhorst electrode m...

Page 7: ...T mode firmly insert the RH T S1 sensor into the top of the meter making sure to align the pin out on the DIN connector 1 From the Main Menu use the key to highlight RH T Then press the SELECT key to...

Page 8: ...RYWALL or SCAN mode screen and an audible alarm will sound if that value is reached Note Default alarm settings are WOOD 15 DRYWALL 1 SCAN 300 MAIN MENU OPTIONS METRIC IMPERIAL This option will toggle...

Page 9: ...ipes or condensation on cold water lines in attic or hollow walls Faulty flashing around windows doors and where porch and dormer roofs meet sidings End grain wood that is not sealed with paint at all...

Page 10: ...o the hole and sealed to the concrete 4 Insert the protective yellow cap into the sleeve and let the hole acclimate for 72h 5 Remove the cap and immediately insert the sensor to full hole depth Wait a...

Page 11: ...ction and climatic conditions Subsurface test Drill two 1 4 holes 3 4 apart and 1 2 to 2 deep Drive the masonry nails into the bottom of the holes and make the tests as described above Nails must not...

Page 12: ...eter scale The meter displays the MC for two seconds USING SCAN MODE You can also take a reference reading on drywall using the meter s SCAN mode To take a reading firmly press the back of the meter o...

Page 13: ...tion of the pinless mode may be adversely affected Under these conditions allow the meter to acclimate to the temperature conditions in which it will be used for minimum of 1 2 hours or as long as pos...

Page 14: ...is not working properly replace the battery with a new one and check the calibration If this does not resolve the problem go to www delmhorst com and follow the instructions under the Product Support...

Page 15: ...ld stay in calibration follow the instructions in the Care of Your Meter section UNDER NO CIRCUMSTANCES SHALL DELMHORST BE LIABLE FOR ANY INCIDENTAL INDIRECT SPECIAL OR CONSEQUENTIAL DAMAGES OF ANY TY...

Page 16: ...20 5 22 GUM RED 7 8 9 10 12 5 14 5 16 5 19 20 5 22 5 24 HEMLOCK WESTERN 7 8 9 10 5 13 15 17 19 20 5 22 23 5 HACKBERRY 7 8 5 9 9 5 12 13 15 17 18 5 20 22 HICKORY 8 8 5 9 10 11 12 5 14 15 5 17 19 20 5...

Page 17: ...4 5 17 19 21 23 5 26 SPF 9 10 11 5 13 15 5 18 20 5 23 25 28 30 SPF COFI 8 9 10 11 13 15 17 19 21 23 25 TEAK 7 8 8 5 9 11 12 14 15 17 18 5 20 VIROLA 6 5 7 8 9 11 12 5 14 16 18 18 5 20 5 WALNUT BLACK 7...

Reviews: