8 Display-Indications
After switching-on the device, 3 pieces of information appear on the upper line of the
display:
Top row left
=> Device status
Possible status indications (S followed by 3 numbers):
First digit 1
airsampler 15 S
First digit 2
airsampler 30 S
Second digit 2
Memory module present (optional)
Third digit
always 0, no function
Top row middle
=> The actual time
Top row right
=> Alert status
Possible alert status indications:
First digit 1
always 1, no function
Second digit 0
Device is inactive
Second digit 1
active with timing, lead time in progress
Second digit 2
active with timing, measuring time in progress
Second digit 3
active with timing, measuring time is expired
Second digit 4
active without timing
Third digit 0
no results
Third digit 1
Voltage drop
Third digit 2
Shut down pressure exceeded
Third digit 3
Over temperature
Third digit 4
Minimum pressure fallen short
Underneath the upper row, the measuring values for volume flow, temperature,
humidity and absolute pressure are displayed
Alert status
Time
Device status
Set point
Volume flow
Temperature
Actual value
volume flow
Absolute pressure
Humidity