Instruction manual 

 

Air sampling device airsampler 15 / 30 S 

 

Page 11 

 

9.1 Possible adjustments / Code chart 

 

Code Function 

Value 

Switch the device active / inactive  

0 = inactive 
1 = active with timing 
2 = active without timing 

10 Volume 

flow 

2 – 15 l/min (airsampler 15 S) 
4 – 30 l/min (airsampler 30 S) 

11 

Measuring time 

5 to 5760 minutes 

12 

Lead time (Time-displaced start) 

5 to 5760 minutes 

20 

Shut down pressure 

50 to 500 mbar 

21 

Minimal pressure 

0 to 100 mbar 

40 

Time hours 

00 to 23 

41 

Time minutes 

00 to59 

42 

Date day 

00 to 31 

43 

Date month 

01 to 12 

44 

Date year 

00 to 99 

51 

Storage interval every X minutes 

1 to 240 minutes 

52 

Number of memory records to be transferred to 
USB memory stick 

0 – 9998 

53 Memory 

delete 

0 = cancel 
1 = delete 

 
 

Summary of Contents for airsampler 15 S

Page 1: ...Instruction manual Air sampling device airsampler 15 S 30 S Manufacturer deconta GmbH Im Geer 20 46419 Isselburg Description Type No airsampler 15 S Type 485 airsampler 30 S Type 486 Serial No...

Page 2: ...3 2 Transport 5 4 Scope of delivery 5 5 Technical description 6 6 Technical data 7 6 1 Pump power airsampler 15 S 7 6 2 Pump power airsampler 30 S 7 6 3 Technical data 7 7 Initial operation 8 8 Displa...

Page 3: ...products guarantee Stability long life and serviceability on site Mechanics with kick Pleasing design The copyright of this instruction manual remains with deconta This manual is intended for assembl...

Page 4: ...ability 2 2 Operations In order to ensure the safety during the operation of the device you absolutely have to have to note the following Do not place in an explosive area Necessary repairs maintenanc...

Page 5: ...handling or carelessness it is common practice to handle the Transport with care 3 2 Transport In order to protect the device during the transport it is located in a transport box with handle Care mus...

Page 6: ...ements Measured are Volume flow temperature relative humidity absolute and barometric pressure The time of measurement lost through power failure is re established A dry contact enables the connection...

Page 7: ...Power connection 230 V Switched power Dry contact 230 V 1 5 A Measurement connections Rods with tube 8 x 1 mm Volume flow 15 S 2 15 l min Volume flow 30 S 4 30 l min max negative pressure 500 mbar Ca...

Page 8: ...pling device airsampler 15 30 S Page 8 7 Initial operation Set up the device Put together the rods and insert them onto the airsampler port Connect the monitor Turn on the device Now it is ready for u...

Page 9: ...First digit 1 always 1 no function Second digit 0 Device is inactive Second digit 1 active with timing lead time in progress Second digit 2 active with timing measuring time in progress Second digit...

Page 10: ...djustments are always carried out according to the same procedure via the keypad Push the key C will appear on the display Enter the code for the relevant adjustment see code chart below Push the key...

Page 11: ...30 l min airsampler 30 S 11 Measuring time 5 to 5760 minutes 12 Lead time Time displaced start 5 to 5760 minutes 20 Shut down pressure 50 to 500 mbar 21 Minimal pressure 0 to 100 mbar 40 Time hours 00...

Page 12: ...10 15 04 8 0 l min 78 mbar 25 C 37 3360 l 60 min 29 04 10 14 04 7 9 l min 61 mbar 25 C 38 2880 l 120 min 29 04 10 13 04 8 0 l min 56 mbar 25 C 37 2400 l 180 min 29 04 10 12 04 8 1 l min 51 mbar 25 C 3...

Page 13: ...mum pressure 0 mbar 29 04 10 16 04 0 l min 0 mbar 25 C 37 3840 l 0 min 29 04 10 15 04 8 0 l min 78 mbar 25 C 37 3360 l 60 min 29 04 10 14 04 7 9 l min 61 mbar 25 C 38 2880 l 120 min 29 04 10 13 04 8 0...

Page 14: ...g Product Negative pressure measuring device Type 485 486 The design of the units complies with EU Machine directive 98 37 EWG the following directives EU Low voltage directive 2006 95 EG EU Directive...

Reviews: