vPad-XSC for Impulse 4000 USER MANUAL
This will result in a typical test report result:
___________________________________________________________
Task:
Charge when Impulse 4000 shows MAXESTART, etc.
Data:
Defibrillator Charge Time
Test:
Energy Limit: 360 J +/- 15 %
Energy: 363.9 J
Maximum Charge Time: 10 s
Result: 21.0 s
Result:
FAIL
___________________________________________________________
3.3.3 Defibrillator Sync Delay (Cardioversion) Test
To perform a defibrillator sync delay test, use a vPL statement of the form
xtest “<
instr
>" | DS<
range
>,<
wave
>,<
targ
>
where:
< instr >
is an instruction to be displayed to the user while the sync test is in progress,
DS
is the command to measure defib sync delay,
< range >
can be “
L
” for low range (up to 50 joules) or “
H
” for high (up to 600 joules),
< wave >
is ECG waveform designator, selected from the ECG table with
*
indicators, and
< targ >
is an optional energy target (in joules) for automatic pass/fail assessment.
Note the comma (“
,
”) separating the required parameters from the optional target energy.
3.3.3.1 Example 10: Measuring Defibrillator Sync Delay
A Checklist may start a defibrillator sync delay test with a target of 20 joules through a vPL
statement similar to the following:
xtest “Set defib to sync mode and discharge now!" | DSH,AF2,20
| 5
For this statement,
DSH
commands
Phase 3
to prepare for a high-range, sync’ed discharge;
AF2
is the waveform designator from the ECG table; and
20
is the optional energy target (J).
Operation/Chapter 3
#
Page 17