vPad-XSC for Impulse 4000 USER MANUAL
1.3 Checklists and vPL
In vPad-Check, equipment maintenance and inspection procedures are called
Checklists
. A Checklist can
contain a sequence of prompts for the operator as well as machine instructions to configure the vPad
safety analyzer and to conduct other tests through a remote instrument. In this case, the remote instrument
is an
Impulse 4000
defibrillator and transcutaneous pacemaker analyzer.
In a Checklist, prompts and machine instructions are written as statements in
vPad Programming
Language
(vPL). A statement generally consists of a specific keyword followed by one or more
parameters. For example, the statement
xopen "Impulse 4000 XSC" | D | 5
initializes a data connection between the Interface and
Impulse 4000
using the keyword “
xopen
”. Other
statements are responsible for configuring and initiating tests through
Impulse 4000
.
Checklists can be written on the Android tablet, or on a personal computer (PC), using a text editor, or
using the Datrend vPad-IDE. Once a new Checklist has been written, it can be imported to vPad-Check
through a USB flash drive, via Bluetooth or a cloud service (Dropbox, Google Drive or OneDrive).
Further information on Checklists and vPL, including a full list of vPL keywords, is provided in the
vPad-
Check Operating Manual
which can be opened with the Datrend Docs app:
.
Overview/Chapter 1
#