vPad-XSC for Impulse 4000 USER MANUAL
Table 1
lists the permitted values for the
“lim”
command. Additionally, any limit type can be set to
“
none
”, which disables automatic assessment for the affected test types. If the Interface receives a limit
configuration command with a disallowed limit type, that particular limit will be disabled.
Note that the units of measurement are assumed when specifying a limit value.
TABLE 1: Allowed Values for Limit Configuration
lim
<
limit_ID
>,
<
limit_type
>,
<
param_1
>,
<
param_2
>
TEST LIMIT
ALLOWABLE VALUES in
bold
defib energy
1
target
0
to
999
%/J
percent
or
units
range
0
to
999
J
0
to
999
J
defib charge
time
2
max
0
to
32000
s
-
defib sync
delay
3
min
0
to
32000
ms
-
max
0
to
32000
ms
-
range
0
to
32000
ms
0
to
32000
ms
pacer pulse
current
4
target
0
to
999
%/mA
percent
or
units
range
0
to
999
mA
0
to
999
mA
pacer pulse
rate
5
target
0
to
999
%/PPM
percent
or
units
pacer pulse
width
6
target
0
to
999
%/ms
percent
or
units
N/A
7
pacer paced
refractory
period
8
min
0
to
32000
ms
-
max
0
to
32000
ms
-
range
0
to
32000
ms
0
to
32000
ms
pacer sensed
refractory
period
9
min
0
to
32000
ms
-
max
0
to
32000
ms
-
range
0
to
32000
ms
0
to
32000
ms
Operation/Chapter 3
#
Page 7