Pressure Module, M-P and Pressure Temp Module, M-PT
Document No. 800 1013-1
17
Modules with signal output connector
•
Preset InvBP and ECG measurement settings:
Invasive Pressures
- P3 Setup - Label - Art
ECG
- ECG1 LEAD -I
Connect a patient simulator to the connector P3 and to the connector ECG on the used
M-NE(12)STPR/M-ESTPR/M-ESTP.
The settings and checks with a Dynatech Nevada medSim 300 Patient Simulator are:
SENSITIVITY switch position: 5 µV/V/mmHg
BP - 1 - WAVE - ART
ECG - BASE - BPM - 160
PACE - WAVE - NSR
Use the channel BP1 for the connector P3.
14.
Connect an oscilloscope between the signal out connector pins 1 (ECG out) and 4 (Ground).
1
2
3
4
Check that an analog signal which corresponds with the ECG waveform on the screen comes
out. The output signal’s ratio to the actual ECG signal should be around 1V/1mV.
Check that a similar output signal is coming out also from the pin 3 (ECG out).
?
15.
Connect the oscilloscope between the signal out connector pins 2 (Pressure out) and 4
(Ground).
Check that an analog signal which corresponds with the InvBP waveform on the screen
comes out. The output signal’s ratio to the actual InvBP signal should be around
1V/100mmHg.
?