Datex-Ohmeda S/5 monitors
Document No. 8001013-1
16
The settings and checks with a Dynatech Nevada medSim 300 Patient Simulator are:
SENSITIVITY switch position: 5 µV/V/mmHg
ECG - BASE - BPM - 60
BP - 2 - WAVE -
ATM
connect a cable from the channel BP3 to the connector P3.
Ccheck that appropriate InvBP waveform is shown and the InvBP value is approximately
25/10 (
±
2 mmHg).
?
Modules with temperature measurement
11.
Check that the “Cable” and “Probe” show OFF for the channels T3 and T4 when no probes are
connected.
Connect a temperature test plug into the connector T3. Check that the “Cable” and “Probe”
for T3 show ON and the corresponding temperature value appears onto the screen.
Perform the same check also for the channel T4.
?
12.
Check the temperature calibrations using temperature test plugs.
If the deviation on a temperature reading on the screen is more then 0.1 °C, calibrate the
temperature channels according to the instructions in the Technical Reference Manual.
?
13.
Activate the temperature test by selecting TEMP TEST from the menu and pressing the
ComWheel twice. When the message “Performing temp test” disappears from the digit field
for T3+T4, check that no error messages appear and “Temp error” shows OFF for both
channels in the menu.
?