APPENDIX A, Service Check Form, M-NMT
A-1(2)
Document No. 800 1018-1
SERVICE CHECK FORM
NeuroMuscular Transmission Module, M-NMT
Customer
Service
Module type
S/N
Service engineer
Date
OK = Test OK
N.A. = Test not applicable
Fail = Test Failed
OK
N.A.
Fail
OK
N.A.
Fail
1.
Internal parts
2.
External parts
3.
Installation
4.
Recognition
5.
Module software
NMT
6.
Communication and
memories
7.
Membrane keys
8.
ElectroSensor
recognition
9.
Stimulus current test
10.
Supramaximal current
< 70 mA
11.
TOF measurement with NMT simulator
T1%
950-1059
T2%
950-1059
T3%
950-1059
T4%
950-1059
Ratio%
950-1059
TOF%
95-105
Count
4
T1%
95-105
12.
Noise
< 50