Datex-Ohmeda S/5 monitors
Document No. 800 1018-1
10
•
Turn the monitor on and wait until the normal monitoring screen appears.
•
Configure the monitor screen so that information regarding the NMT measurement is shown,
for example:
Monitor Setup
- Screen 1 Setup - Digit Fields - Field 4 - Nmt
•
Preset the NMT measurement settings:
Others
- NMT - Stimulus Mode - TOF
Set Cycle Time - 10 sec
NMT Setup - Current - S(70 mA)
Pulse Width - 200
µµµµ
S
Stim. Beep Volume - 2
3.
Plug in the module. Check that it goes in smoothly and locks up properly
?
4.
Check that the module is recognized, i.e. the NMT header with related information appear in
the chosen digit field.
?
5.
Enter the service menu:
Monitor Setup
- Install/Service (password 16-4-34) - Service (password 26-23-8
)
Take down the information regarding NMT software by selecting SCROLL VERS and turning
the ComWheel.
?
6.
Enter the NMT module service menu:
Parameters - NMT
Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not increasing
faster than by 50 per second. Check that the module’s memories have passed the internal
memory test, i.e. “RAM”, “ROM” and “EEPROM” all state OK.
?
7.
Check the front panel START-UP and STOP/CONTINUE membrane keys.
Press each key for at least one second and check that the key being pressed is identified,
i.e. the color under the corresponding text on the menu turns blue.
?