NeuroMuscular Transmission Module, M-NMT
Document No. 800 1018-1
13
19.
Perform an electrical safety check and a leakage current test.
?
20.
Check that the module functions normally after performing the electrical safety check.
?
21.
Clean the module with suitable detergent.
?
•
Fill in all necessary documents.
3.3
Disassembly and reassembly
Disassemble the NeuroMuscular Transmission Module, M-NMT, in the following way. See the
exploded view of the module.
1.
Remove the two screws from the back of the module.
2.
Pull the module box slowly rearward and detach it from main body. Be careful with the loose
latch and spring locking pin.
3.
Detach the NMT board by removing the two screws located near the front panel frame,
disconnect the cables and pull out the front panel frame.
To reassemble the module, reverse the order of the disassembly steps.
CAUTION
When reassembling the module, make sure that the cables are reconnected
properly.