NeuroMuscular Transmission Module, M-NMT
Document No. 800 1018-1
17
5.1
NMT menu
Start Curr. test Start Curr. test
is a test where the module itself checks the difference between a given current and
the measured current. A 3 k
Ω
resistance should be connected between the stimulus electrodes
before starting the test. All the currents checked will be displayed on the service data screen. After
the current value, the status of the test is shown. If the test is not passed, send the module back to
the factory for calibration.
T1%/ref
T1% / ref
gives a selection for changing settings related to the M-NMT MechanoSensor. A setting
of
3
should always be used for normal monitoring. The settings
1
and
2
are for research purposes
only.
Record Data
Record Data
prints out the displayed service data and circuit board information (id., serial number,
and software id.) onto the Recorder Module, M-REC.
Service Data
Cable
shows the type of cable being used.
Cable check value
shows the bit amount. From the following chart you can check the value for
each cable:
EMG
-100...+100
Cable OFF
>1950
Piezo
900...1100
Regional block
600....800
Module Keys
checks the function of the module keys. A blue background appears at the back of
the text when a key is pressed for more than one second.
Stimulus mode
shows the selected stimulus. Stimulus mode can be changed using the NMT Setup
menu.
Measurement
indicates ON/OFF.
T1%, T2%
shows the measured response. A value of 1000 corresponds to 100%.